Unassignment
7.
A "Data Collection Complete" tag will appear in the general tab under the "Subject Details" page.
8.
Collect the CPW01, charging dock, and all accessories. Clean the CPW01 with an alcohol-based solution (wipes or
swabs) and store all equipment until the end of the study.
CPW01 devices from ActiGraph are approved to be cleaned with a non-abrasive cloth and any alcohol-based
solution. For optimal results, apply isopropyl alcohol to the cloth, clean the device for 15-20 seconds, and let air
dry. The ActiGraph CPW01 devices are not to be sterilized because damage may occur. Although the CPW01 devices
from ActiGraph are not required to be cleaned on any predetermined schedule, it is advised that the devices be
cleaned after each use by a user.
Cleaning the CPW01 multiple times during its expected service life will not cause a loss of basic safety or essential
performance.
Please note: The CentrePoint Insight Watch (CPW01) and accessories should be cleaned with 70%
Isopropyl alcohol between each use. Allow 10 minutes to air dry prior to re-use.
CENTREPOINT INSIGHT WATCH | E.200.6002 | RELEASED: 01/09/2023 | REV: 7
|
Continued
19
THEACTIGRAPH.COM