DDI PRODIGY AUDIO 51720 User Manual page 16

Talking, no-coding required!, talking no-coding required!
Table of Contents

Advertisement

Available languages

Available languages

P R O D I G Y
B)
To Calibrate When the Meter
Has a Test Strip Inserted
Step 1
– Insert the Test Strip
Start with the meter off. Insert a test
strip into the test slot. LCD displays
"CH" and "
temperature next, and finally a
flashing code number.
Step 2
– Match the Code Number
Compare the flashing number with
the code number on the test strip
vial. If the two numbers match, press
the M button or wait 5 seconds to fix
the code. Then " " appears, indicat-
ing that you can directly begin
testing. If they do not match, proceed
to Step 3.
Step 3
– Select a Correct Code
While the code number is still flash-
ing, press the Code button to
advance one number. When the
correct code appears, press the M
button or wait 5 seconds to fix the
code. Then " " appears. You can
begin testing. If you would like to turn
off the meter, remove the test strip or
wait 2 minutes for the meter to auto-
matically shut down.
Caution: If the code number
displayed on the meter does not
match the number printed on the
vial, test results may be inaccurate.
Diagnostic Devices, Inc.
16
®
Blood Glucose Monitoring Systems – User's Manual
" first, the ambient
www.prodigymeter.com
Figure 10
Figure 12
1- 800-243-2636
Figure 11