Screw-in ⁄ immersion temperature radio sensors

Advertisement

®
KYMASGARD
KYMASGARD
®
Screw-in ⁄ immersion temperature radio sensors
APPLICATION:
Built-in temperature sensors are electric contact thermometers for installation in pipes, tanks, and ves-
sels  for  measuring  temperatures  of  liquids  and  gases.  Screw-in  resistance  thermometers  with  plastic 
enclosure made of impact-resistant plastic and straight protective tube. For aggressive media, stainless 
steel  immersion  sleeves  are  to  be  used.  Applications  in  piping,  in  heating  technology,  in  storage  tanks, 
district heating compact stations, warm and cold water systems, in oil and lubricant circulating systems, 
in mechanical, apparatus and plant engineering, as well as in the entire industrial sector. 
TECHNICAL DATA:
Measuring range: .........................................-  2 0  . ..+  8 0  ° C 
Sensors ⁄ output:............................................radio signal
Process connection:  . ..................................screwed socked with ½" pipe thread  
Immersion sleeve:  . ........................................ETF  1 -FSE: brass, nickel-plated,  
Protective tube: .............................................stainless steel, 1.4571, V    4 A, Ø  =   6   m m
Connecting head: ..........................................plastic, material polyamide, 30  %  glass-globe-reinforced,   
Dimensions:   . ....................................................108  x   7 0  x   7 3.5    m m 
Max. pressure:  . ...............................................ETF  1 -FSE: brass immersion sleeves 10  b ar 
Humidity:  . ............................................................<    9 5  %  r.  H .
Protection class:  . ..........................................III   ( according to EN 60  7 30)
Protection type:  . ............................................IP 65   ( according to IEC 529)
Standards:    . ........................................................CE conformity, electromagnetic compatibility according  
Radio receiver  :  ...............................................see end of chapter 
Basic device:  . ....................................................see pages 28  -   2 9
Dimensional drawing 
Type ⁄ WG 3 - EL 
ETF 1 - FSE 
ETF 2 - FSE 
Optional: 
Please indicate with your order: 
 
ETF 1-  F SE 
including brass immersion 
sleeve, nickel-plated
ETF 2-  F SE 
including stainless steel
immersion sleeve
½" pipe thread wrench size 22 mm, Ø  =   8   m m  
ETF  2 -FSE: stainless steel, 1.4571, V4A,  
½" pipe thread, wrench size 27 mm, Ø  =   8   m m
colour pure white (similar RAL  9 010) 
ambient temperature -    2 0  . ..+100  ° C
ETF  2 -FSE: stainless steel immersion sleeves 40  b ar 
to EN 61  3 26  +   A 1  +   A 2, EMC directive 89 ⁄ 336 ⁄ EWG
50 mm 
100 mm 
• 
• 
• 
• 
Solar module (enclosure with solar cell) 
Type and length of immersion sleeve   
e.g. ETF 1-FSE, 150 mm;  ETF 2-FSE, 250 mm 
ETF x - FSE
150 mm 
200 mm 
• 
• 
• 
• 
250 mm 
300 mm 
• 
• 
• 
  
ETF x - FSE 
(standard)
ETF x - FSE 
(solar cell)
400 mm