Operator ID (OID) configuration:
Tap the Operator ID request of choice on the screen:
• No: The meter does not request operator identification when a patient
or QC test is initiated.
• Optional: The meter requests an operator identification when a patient
or QC test is initiated and if the operator has not previously logged on,
but this request may be skipped.
• Required: The meter forces operator identification when a patient or
QC test is initiated and if the operator has not previously logged on.
The selected option is highlighted (bold type with a check mark).
This setting is optional by default.
Tap Confirm to save this setting or tap Back to return to the previous
screen without saving the changes.
Patient ID (PID) configuration:
Tap the Patient ID request of choice on the screen:
• No: The meter does not request patient identification when a patient's
test is initiated.
• Optional: The meter requests patient identification when a patient's
test is initiated but this request may be skipped.
• Required: The meter forces patient identification when a patient's test
is initiated.
The selected option is highlighted (bold type with a check mark).
This setting is optional by default.
Tap Confirm to save this setting or tap Back to return to the previous
screen without saving the changes.
3.4 EASYCONTROL SETTINGS
The EasyControl Settings menu contains options for its configuration.
These settings allow the optional liquid control tests to be run on a desired time basis.
This option is inactive by default.
These settings must be configured by the administrator (the administrator must be logged on).
Lockout, Frequency and Lock Mode
Tap Settings from the main menu to define the EasyControl Settings.
Then tap EasyControl Settings.
Slide the ON/OFF button to activate the EasyControl lockout.
Then set the Frequency for the microINR EasyControl scheme. Tap the
selected frequency on the screen and a new window will open:
• No: the meter will only be locked in case an incorrect EasyControl result
is obtained.
• Daily
• Weekly
• Monthly
• Number of tests: from 1 to 2,000 tests.
Various options can be configured according to the following description
whenever the lockout option is active and the microINR EasyControl test
is not conducted properly or the result is outside the acceptance range:
16