OfiTE 120-75 Instruction Manual

OfiTE 120-75 Instruction Manual

Atmospheric consistometer

Advertisement

Quick Links

Atmospheric Consistometer
#120-75 (115V)
#120-75-1 (220V)
Instruction Manual
Updated 3/6/2023
Ver. 8
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
Copyright OFITE 2015
©

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 120-75 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for OfiTE 120-75

  • Page 1 Atmospheric Consistometer #120-75 (115V) #120-75-1 (220V) Instruction Manual Updated 3/6/2023 Ver. 8 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com Copyright OFITE 2015 ©...
  • Page 2: Table Of Contents

    Operation..................8 Calibration ..................9 Torque Head ................9 Temperature ................10 Motor Speed ................10 Timer ..................10 Test Setup ..................11 Maintenance ..................13 Warranty and Return Policy ............15 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 3: Intro

    Crated Weight: 200 lb (90.7 kg) Requirements Water Supply for Cooling Water Drain 220 Volt, 50/60 Hz, 2.2 KVA Power Source 120 Volt, 50/60 Hz, 4.4 KVA Power Source OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 4: Components

    #121-010-1 Heater #122-073-1 Fuse, 3 Amp, 5 mm × 20 mm #122-075 Fuse, 7 Amp, 5 mm × 20 mm #220-15A-EURO AC Power Cord OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 5 Fuse, 10 Amp, 5 mm × 20 mm, Qty: 3 #152-55 Bath/Circulator, Refrigerated and Heated, 115 VAC, 60 Hz #152-55-1 Bath/Circulator, Refrigerated and Heated, 230 VAC, 50 Hz OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 6: Setup

    To increase the rotational speed, turn the potentiometer speed control clockwise. Turn it counterclockwise to reduce the speed. Potentiometer Speed Control Coolant Supply Coolant Drain Tank Drain OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 7: Loading The Test Cells

    5. Lower the test cell into the bath and lock it in place with the locking pins. Torque Shaft Locking Pin Locking Pin Locking Pin OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 8: On-Screen Display

    Press Stop Cooling to close the valve and stop the flow of water. The Start Heating button turns on the heater. Press the Stop Heating button to turn the heater off. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 9: Operation

    Thoroughly clean the containers with soap and water. Be sure to remove any residual cement. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 10: Calibration

    When adding weights, remember that the hook weighs 50 grams. Therefore, to test the potentiometer at 200 g, you only need to add 150 g to the hook. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 11: Temperature

    The timer should be accurate to within one second per 12 minutes. To verify the timer, start the timer at the same time as a certified timer and compare the two readings after 12 minutes. Timer OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 12: Test Setup

    3. To change the name of the profile, touch the profile name above the step list. A keyboard will appear. Type the profile name and touch “Enter”. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 13 The timers on the “Real-Time Data” screen will be disabled while a test is running. 8. To stop the test, touch “Test Setup” and then “Stop Test”. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 14: Maintenance

    Pull the motor back only enough to prevent belt slippage. Allow approximately ¾" to 1" slack in the timing belt to prevent excessive side load to the bearings. Belt Bearing Agitator Shaft Rotor OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 15 Remove all of the set screws in the front and rear side of the drive housing. b. Push the bearing up and out from below. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 16: Warranty And Return Policy

    In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties...

This manual is also suitable for:

120-75-1

Table of Contents