ES measurement is not appropriate. Only trends in ES should be used in making treatment decisions. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
0.061" ± 0.001" (1.55 ± 0.03 mm) at 72°F (22°C) The probe, standards, and batteries are stored inside the ES Meter case. Twist the lock to open the door and access these items. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
5%. If they differ by more than 5%, the meter or electrode should be checked. 8. Record the average of the two ES measurements on the Drilling Mud Report Form. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
The ES meter must be periodically checked by a qualified technician to ensure that all units meet specifications. OFITE recommends returning the ES Meter once a year for a thorough calibration check. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...