OfiTE 1100 Instruction Manual

OfiTE 1100 Instruction Manual

Pressurized viscometer with orcada software control system

Advertisement

Quick Links

Model 1100 Pressurized Viscometer
with ORCADA
Software Control System
®
#130-81-C - With Laptop PC and Padded Carrying Case, 115-Volt
#130-81-1-C - With Laptop PC and Padded Carrying Case, 230-Volt
Instruction Manual
Updated 4/25/2019
Ver. 10
OFI Testing Equipment, Inc.
11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A.
Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com
Copyright OFITE 2015
©

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 1100 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for OfiTE 1100

  • Page 1 #130-81-1-C - With Laptop PC and Padded Carrying Case, 230-Volt Instruction Manual Updated 4/25/2019 Ver. 10 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. · Houston, Texas · 77065 · U.S.A. Tele: 832.320.7300 · Fax: 713.880.9886 · www.ofite.com Copyright OFITE 2015 ©...
  • Page 2: Table Of Contents

    Main Bearing Replacement ............39 Rotor Nut ...................42 Fuse Replacement ..............43 Appendix ..................44 Diagram ..................44 Transducer Linearization ............46 Driver Install ................49 Warranty and Return Policy ............52 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 3: Intro

    AC Power Cord; 3-Conductor International (Continental Euro- pean) #153-00 Bottle Brush #153-55 Silicone Stopcock Grease; Dow Corning; 150g Tube #153-67 60cc Disposable Syringe #170-17 Valve Stem O-ring, -008, Viton 75D OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 4 Bath/Circulator; Caron Model 2050W; Refrigerated and Heat- ed; 115V #152-55-1 Bath/Circulator; Caron Model 2050W; Refrigerated and Heat- ed; 230V #165-44-2 High-Temperature Thread Lubricant; 7 gram Pouch OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 5: Features

    ® temperature without the risks of hot oil, such as spilling, splashing or flashing. *U.S. Patent Number 8,739,609 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 6: Specifications

    * Lower shear rates available on special order. ** At 1,000 RPM *** At .01 RPM When using the 130-81-079 spring use a maximum speed of 300 RPM. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 7 To convert from units on left side to units on top, multiply by factor @ intercept. Dyne/cm lb/100ft lb/ft Dyne/cm 0.2084 0.002084 0.1957 2.084 0.02084 1.957 lb/100ft 4.788 0.4788 0.01 0.939 lb/ft 478.8 47.88 93.9 5.107 0.5107 1.065 0.01065 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 8: Setup

    AC power cord plugged in overnight. 3. Turn the monitor and computer on. After the computer has booted up completely, check for the ORCADA icon on the desktop. ® OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 9: Viscometer

    5. Make sure the “Power” switch is in the off position. Plug the power cord into an AC power source. Serial Port Ethernet Port Nitrogen Power Drain Water In OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 10: System Test

    Model 1100 Viscometer, however other combinations are available. See page 5 for more information. Sample Cup Nut Sample Cup Valve Stem Heat Bath OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 11 8. It is recommended that a Calibration procedure be run due to the possibility of mishandling during shipment. Refer to the “Calibration” section on page 16 for detailed instructions on calibrating the unit. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 12: Preparing For A Test

    The alarm light will turn green when the appropriate pressure is applied to the sample. Refer to the chart below for the minimum pressure requirements. 6. Once the pressure is set and the heater is in place, the Model 1100 is ready to run a test. Temperature Pressure Ambient - 200°F...
  • Page 13: Software Start

    “Cement” - This button performs a standard cement test based on API specifications. “Mud” - This button performs a standard mud test based on API Specifications. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 14 If you choose to “Override”, the time remaining in the conditioning cycle will resume at the point where the test was paused. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 15 Enter a value in the “Temp” field and check the “PreHeat” box to preheat the bath prior to heating the sample. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 16 The drop-down box next to the “Shear Stress” field changes the units of the Shear Stress variable. The units can be set to: Dyne/cm , lb/100ft , Pa., Dial Reading, lb/ft OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 17: Calibration

    When using the 130-81-079 spring, use 200 cP fluid with a B5 bob and 100 cP fluid with a B1 bob. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 18 8. When the calibration is complete, check the r^2C value to make sure it is greater than .9990. If it is, click the “OK” button and enter a title for the calibration. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 19 “Ref Viscosity” - the viscosity of the sample as calculated by the software “Ref Shear Stress” - the shear stress as calculated by the software OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 20: Zeroing The Unit

    Observe the signal value after 15 seconds to make sure the signal does not drift out of range. 7. Tighten the thermocouple compression fitting. Thermocouple Compression Transducer Cap Fitting Set Screw OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 21: Fluid Manager

    4. Once you have entered all the batches, click “OK” to close the Fluid Manager and return to the Calibration screen. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 22: Operation

    Allow the sample temperature to drop below 120°F (49°C) before reducing the pressure. Make sure the pressure reads 0 before removing the sample cup. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 23: Auto Mode

    Allow the sample temperature to drop below 120°F (49°C) before reducing the pressure. Make sure the pressure reads 0 before removing the sample cup. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 24: Software

    “Logo Files” - Specify a logo file (.jpg format) to display on the graph at the end of the test. “Time Format” - Choose a time format for all of the ORCADA software. ® OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 25: Save Rate Settings

    “Shear Stress Tare” - change in measured shear stress “Pressure” - measured pressure “Pressure SP” - pressure setpoint “pH” - Not available on the Model 1100. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 26: Test Builder

    “Heat RPM” - the rotational speed during the heating period c. “Temp” - the temperature setpoint. Enter any letter in this field to indicate “Ambient”. Available Tests Records Test Detail OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 27 “Stir Parameters” - These parameters only apply if the test includes a gel time. “Stir Prior to Gel” - Check this box to stir the sample before calculating gel strength. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 28 Cursor” button moves the cursor to the center of the graph. Click and drag the cursor across the graph to see more detailed information. Refer to the “Test Detail” field below the graph for more information. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 29: Select Analysis Outputs

    To remove a data point from the chart, simply uncheck the box next to it. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 30: Saved Test Data

    “Analyzed Data” tab shows the results of the calculations based on the Analysis Model. The “Test Setup” tab shows the parameters of the test. Raw Data OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 31 “Experiment Explorer” - Click this button to choose another set of test data to review. “OK” - Click this button to return to the main screen. Raw Data Test Setup OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 32: Calibration History

    6. When you are finished reviewing the calibration history, click the “OK” button to return to the main screen. Calibration History Explorer Calibration History OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 33: Import/Export

    7. Click the “Import” button. The new tests will appear in the list in the upper left-hand corner of the Main Screen. Export Import OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 34: Disassembly

    8. Open the panel on the right-hand side of the unit cabinet. 9. Inside the cabinet, locate one of the six holes on the rotor tube. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 35 13. Reinstall each of the removed parts. Remember to hold the rotor tube in place while tightening the expansion joint. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 36: Maintenance

    Allen bolts into the viscometer. Remove the transducer cap and set it aside. Be very careful to keep the thermocouple straight. A bent thermocouple will cause inaccurate readings. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 37 10. Slide the lower bearing and bearing retainer off the bob shaft. 11. Slide the bob shaft up out of the transducer body. Bob Shaft Bob Shaft Lower Bob Shaft Bearing Transducer Body OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 38 Then turn the bob shaft back slightly and tighten the set screw on the torsion spring to secure it to the bob shaft. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 39 950 and 1000. See page 19. 26. Run the TransCal software as instructed on page 46. 27. Re-calibrate the unit before performing another test. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 40: Main Bearing Replacement

    Do not touch the new bearings with your bare hands. Oils from your skin can damage it. Main Body Rotor Cap Nut Main Bearing OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 41 Do not touch the new bearings with your bare hands. Oils from your skin can damage it. Lower Bearing Main Seal Retaining Ring Seal Nut OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 42 Tighten the Allen bolts on the motor to secure it in place. 13. Reassemble the head assembly as described in step 19 - 26 on page OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 43: Rotor Nut

    4. Put the rotor tube back in place and secure it with the rotor nut. Make sure the nut is tight. 5. Reassemble the head assembly as described in step 19 - 26 on page 38. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 44: Fuse Replacement

    4 amp 2 amp Main Power Supply 2 amp 1 amp SSR - Heater 3 amp 3 amp SSR - Ram 3 amp 3 amp OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 45: Appendix

    Appendix Refer to the next page for a list of parts that correspond to the labels on this diagram. Diagram OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 46 Hex Socket Head Cap Screw; 5/16-24 UNF - 1 130-81-060 R1 Rotor Assembly 142-54 O-ring for Valve Stem Port 170-17 O-ring for Valve Stem OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 47: Transducer Linearization

    “Name” field. 3. Set the following port settings: Bits per second: 38400 Data bits: 8 Parity: None Stop bits: 1 Flow control: None OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 48 8. Ignore the instruction on the screen. Slowly turn the bob clockwise until it reaches the other stop. 9. Type N and press Enter. The software will show details of the calibration. Close Hyperterminal. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 49 Close the TransCal software and open the ORCADA software. ® 14. Calibrate the viscometer through the ORCADA software before running ® any other tests. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 50: Driver Install

    3. The FTDI CDM Driver will open to allow you to extract the driver and launch the installer. Click “Extract”. 4. A welcome window will open. Click “Next”. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 51 Click the Start Menu. Open the “Control Panel”. Click the “Hardware and Sound” link. Click on the Device Manager link in the “Devices and Printers” options. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 52 This is the COM Port that the driver has been assigned to. 10. Open the Options screen in the ORCADA software. Choose the COM ® port listed in Device Manager. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
  • Page 53: Warranty And Return Policy

    In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied,...

This manual is also suitable for:

130-81-c130-81-1-c

Table of Contents