The resulting pressure increase is then measured and compared to the calibration curve to determine the total weight of CaCO in the test sample. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Do not take internally. In the event of accidental exposure, get medical attention immediately. Refer to the Material Safety Data Sheet (MSDS) for more information on Hydrochloric Acid OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
Begin by carefully unpacking the equipment. Inspect each piece to ensure everything is clean and in good operating condition. Inspect the cell cap o-ring for cracks or wear. Gauge Cell Cap Bleed Valve O-ring OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
4. Turn the reaction cell back to vertical and start timing the test. This will start the reaction between the HCl and the CaCO OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
.05 × 100 = 5. This number is the slope or average slope for the equipment. It is a function of the volume of the reaction cell. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
2. Check the reaction cell for contaminants. Be sure the cell is clean and dry. 3. Check for impurities in reagents. Moisture in CaCO will result in low- pressure readings. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com...
In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied,...