Product Feature - Haier HXC-106 Operation Manual

Blood bank refrigerator
Table of Contents

Advertisement

Product Feature

Haier blood bank refrigerators are applicable for the storage of whole blood, medicine, biological
products and other laboratory products that need to be stored at a temperature of 4℃ . Haier blood
bank refrigerators are widely installed in blood stations, hospitals, colleges, universities, research
quarantine stations and other research and clinical facilities.
1.Temperature control
•Computer control, digital display of temperature, temperature display accuracy as 0.1°C and
temperature range as 4°C ± 1°C (temperature range of HXC-1308 and HXC-1308B as 4°C ± 1.5°C).
2.Security system
•Multi-fault alarms (high and low temperature alarm, low battery alarm, power failure alarm, sensor
fault alarm and door opening alarm).
•Two alarm methods (buzzing alarm, fl ashing alarm).
•Safe grounding of all individual components.
•Standard remote alarm interface, easy to use and reliable.
3.Refrigeration system
•Optimized refrigeration system off ers eff ective and effi cient performance.
4.User-friendly Design
•User-friendly design, computer control, smart and carefree, adjustment not required.
•High-performance thermal insulationensures super insulation eff ect.
•Multiple condensation-proof technologies applied on glass doors.
•Automatic evaporation of condensate water, user-friendly design of safety door latch, etc.
Note: We apologize that the Haier blood bank refrigerator provided to you may be a little
diff erent from the fi gure in the manual due to product improvement. The manual may be
subject to changes without further notice.
6
Name
Model
HXC-608
Exterior dimensions
(W x D x H) (mm)
Interior dimensions
(W x D x H) (mm)
Eff ective volume
Door
Insulation
Compressor
Shelves / Drawer
6 shelves / 24 baskets
Load
Refrigerating Method
Exterior / Interior
Condenser/Evaporater
Temperature Control
Refrigerant
Interior light
Net weight
Temperature range
Voltage
Rated power / current
Noise Level
Foaming Agent
Climate Type
Anti-shock Safety
Classifi cation
Power connection type
USB function / temperature
recorder
High-low temperature alarm, sensor fault alarm, power failure alarm,
Alarm system
Battery duration for alarm
system
Rechargeable battery
Note: Climate Type ST means the temperature is 18 - 38℃ ,the relative humidity is up to 85%.Due
to continuous improvements, technical information on your refrigerator may be diff erent from herein
published.
Blood Bank Refrigerator
HXC-608B
780x840x1945
680x640x1400
608L
Glass door with electric heat
CFC-Free foamed-in-place urethane
High Quality Hermetically sealed compressor
6 Drawers
Max:30kg per shelf
Forced convection cooling
Scratch resistant painted steel / Stainless steel
Fin tube type
Microprocessor controlled
R134a 420g
LED 3W
204kg
211kg
4±1℃
220V-240V~/50Hz/60Hz
490W/3.5A
43dB(A)
HFO-1233zde
ST
I
Y
Optional / Standard
low battery alarm and door opening alarm
48 hours (when battery is fully charged)