Internal Hardware Servicing; Trouble Shooting Guide; Troubleshooting - Delfi PTS ii Operator's Maintenance Manual

Personalized tourniquet system
Hide thumbs Also See for PTS ii:
Table of Contents

Advertisement

MAINTENANCE

INTERNAL HARDWARE SERVICING

The PTS ii is designed with self-test and self-monitoring features to warn of failures. Refer to Alarm Conditions under
Section 2 Operating Instructions. Although very unlikely, modes of failure may also occur that cause erratic operation
and/or illegible displays and may or may not trigger alarms. If the maintenance, calibration, and troubleshooting
procedures do not restore normal operation, contact Delfi for service advice.
CAUTION:
Do not attempt to disassemble or open the enclosure of your PTS ii. The PTS ii is not designed to be
disassembled and serviced by anyone other than Delfi staff. Disassembly and attempted service by anyone other than
Delfi staff poses a risk of electric shock, damage to the unit, and injury to the patient and will void all warranties. Internal
parts in the PTS ii can only be serviced at the factory by Delfi staff. Please contact Delfi if you have problems with your
PTS ii that cannot be resolved by following the maintenance and troubleshooting procedures described above

TROUBLE SHOOTING GUIDE

To aid in unscheduled maintenance, the TROUBLESHOOTING table below delineates a number of possible malfunctions
that could occur with the unit. The most likely causes are shown for each symptom. While it is not practical to enumerate
every conceivable malfunction and all possible causes, the table will assist in isolating the most common problems.

TROUBLESHOOTING

Malfunction
Unit does not turn ON (with no AC
power connected)
Unit does not turn ON (with AC
power connected)
Cuff does not inflate
Cuff does not deflate
52
MAINTENANCE
Possible Causes
Battery pack not charged
Battery pack disconnected
Internal hardware failure. Defective
AC adapter/cord assembly
INFLATE button not touched
Hose kinked or blocked
Internal hardware failure
DEFLATE slide not engage
Hose kinked or blocked
Internal hardware failure
Corrective Actions
Plug in to AC power and allow
battery to charge. Attempt to turn
unit ON with AC power. Refer to
"BATTERY FAILURE" alarm below
Remove battery cover and ensure
battery pack is securely plugged in
Contact Delfi
Touch the INFLATE button
Unkink hose or disconnect cuff from
hose.
Contact Delfi
Engage the DEFLATE slider. Refer to
Operation Step 7 under Section 2
Operating Instructions
Unkink hose or disconnect cuff from
hose. Ensure complete cuff
deflation to clear CUFF NOT
3