OPERATING INSTRUCTIONS
2
Initiating a Leak Test:
•
Press the INFLATE button to initiate a leak test
•
During a leak test, the cuff will inflate to approximately
250 mmHg and a 30 second leak test will commence. At
the end of the test, the cuff will automatically deflate,
and the test result will be displayed
WARNING:
Do not start a leak test (press the INFLATE button)
if the cuff is applied to a patient
•
To stop the leak test, touch the STOP button
3a
Leak Test Results (Passed):
•
If the leak test has passed, a "LEAK TEST PASSED"
notification will be visible
•
To initiate a new leak test, touch the RETRY button
•
Touch the BACK button to exit the leak test and return to
the Main Screen as necessary
31
OPERATING INSTRUCTIONS