Appendix A. Fault Finding
Error Message
Probable cause
!019 Timed Out
Key not pressed in the specified time
period (usually 20 seconds) normally
happens with inactive commands.
!020 No
Change of measurement speed, see
Frequency
Section 4.6.3 on page 19.
!021 Bad
Error occurred in memory check sum.
Checksum
!022 Bad
Erroneous character detected in the
Message
middle of a message.
!023 Bad Cal
Divide by zero error occurred in
Pres
calibration calculation.
34 | 8000/8100/8200 Series Instruction Manual-English
Table A-1: Fault Finding
Action
Repeat command, entering response in
less than 20 seconds between each
character and send carriage return within
the specified time.
Wait for 2 seconds, if fault remains return
to GE for repair or replacement.
Power cycle sensor, if fault remains return
to GE for repair or replacement.
Check and re-send message.
Check data and repeat calibration.