Chapter 1. Installation
Male threaded pressure connectors must not be sealed or constrained against the face at the
base of the thread. The forward cone or flat face should always be used as indicated below.
Depth versions should not be used at hyperbaric pressures above 70 bar (approximately 700
m of water).
1.3 Media Compatibility
The media compatibility of the TERPS sensors is shown in Table 1-1.
Product
81XX
80XX
82XX
83XX
Note: Fluid classification complies with Regulation (EC) No 1272/2008. Statements
comply with Pressure Equipment Directive 2014/68/EU.
1.4 Pressure Containment
The pressure containment of the TERPS sensors is shown in Table 1-2.
Product
DPS 81XX
DPS 80XX
2 | 8000/8100/8200 Series Instruction Manual-English
Table 1-1: Media Compatibility
Pressure Range
0 to 3.5 bar
0 to 70 bar
0 to 70 bar
Table 1-2: Pressure Containment
Pressure Range
0 to 3.5 bar
0 to 7 bar
>7 to 70 bar
Media Compatibility
Non-condensing dry gases compatible with silicon,
silicon dioxide, RTV adhesive, stainless steel 316L
and glass.
Fluids compatible with Stainless Steel 316L and
Hastelloy C276.
Fluids compatible with Hastelloy C276.
Pressure Containment
7 bar maximum
70 bar maximum
200 bar maximum
© 2010 Baker Hughes, a GE company – All rights reserved.