Qsep
Bio-Fragment Analyzer:
1
Page: 1/3
1. Ensure the protecting foam is removed (Fig. 1 & 2).
*Scan the QR code for instruction video.
Fig. 1
Fig. 2
Qsep
2. Plug the power cord (labeled
cable (optional).
3. Connect the air tube.
Qsep
4. Switch on
, the LED lights up (multiple color
1
light) (Fig. 3). Wait 25 seconds until the instrument
and Wi-Fi initialized, and the green LED will start
flashing (Fig. 4).
*DO NOT switch on the instrument immediately
after powering off. Wait at least 5 seconds.
Blue
Green
Fig. 3
Fig. 4
*The green flashing LED shows the instrument is
ready to connect. If not, please repeat step 4.
Operation Quick Start
Ethernet cable connection:
refer to step 5 on Page 2
Ethernet cable
) and Ethernet
1
Wi-Fi connection: follow the steps below to
connect with
-
Check if the AP source is available for Wi-Fi network
on the computer (Fig. 5).
Fig. 5
Qsep
1
Document No.: F0016
Ver.: ENG-F
-
Find the SSID named with
"Qsep1+instrument ID
(16 digits) "
and click "Connect" (Fig. 6).
-
Enter the password:
The password will show in the sticker which is
behind the instrument. It is the same as the
8 digits
of instrument serial number (same as SSID
name) without the prefix "Qsep1" (Fig. 7).
00923718
Fig. 7
Qsep
*If the computer cannot find the
Qsep
the distance between
and the computer
1
within 10 meters and start from step 4 again.
*If the computer cannot connect with
-
Check the Wi-Fi IP assignment is "Automatic
(DHCP)"
-
Initiate the software as administrator
Q-Analyzer
-
Allow
pass through firewall or turn
off the firewall
Fig. 6
middle
SSID, keep
1
Qsep
,
1