Amsco Evolution HC-600 Operator's Manual

Amsco Evolution HC-600 Operator's Manual

Medium steam sterilizers
Hide thumbs Also See for Evolution HC-600:
Table of Contents

Advertisement

Quick Links

OPERATOR MANUAL
®
AMSCO
Evolution
Medium Steam Sterilizers
RevAC
P129390-106

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Evolution HC-600 and is the answer not in the manual?

Questions and answers

Summary of Contents for Amsco Evolution HC-600

  • Page 1 OPERATOR MANUAL ™ ® AMSCO Evolution Medium Steam Sterilizers RevAC P129390-106...
  • Page 3 Evolution™ Medium Steam Sterilizer is intended for shipment to locations outside the United States of America. The Amsco Evolution medium steam sterilizer is designed for sterilization of heat- and moisture-stable materials used in healthcare facilities and is in a prevacuum configuration: Prevacuum –...
  • Page 4 Table 1. Prevacuum Sterilizer Cycle Values Validation Cycle Sterilize Temp. Sterilize Time Dry Time Standard PREVAC 134°C (273°F) 3 Min., 30 Sec. 30 Min. or 20 Min. EN 285:2006 PREVAC 121°C (250°F) 30 Min. 30 Min. or 20 Min. EN 285:2006 LIQUID 121°C (250°F) 45 Min.
  • Page 5 Manufactured by: The base language of this document is ENGLISH. Any translations must be made from STERIS Mexico, S. de R.L. de C.V. the base language document. Avenida Avante 790 Parque Industrial Guadalupe Guadalupe, Nuevo Leon, Mexico C.P. 67190 2797 Sales and Service: STERIS Corporation 5960 Heisley Road...
  • Page 6 P129390-106 Operator Manual Introduction...
  • Page 7: Table Of Contents

    TABLE OF CONTENTS Section Number Description Page Safety Precautions ........................1-1 Symbol Definitions ........................2-1 Installation Verification ......................... 3-1 Installation Checklist ..........................3-1 3.1.1 Service Clearance........................3-1 3.1.2 Plumbing Services ........................3-1 3.1.3 Electrical Service.......................... 3-2 3.1.4 Sterilizer Final Check ........................3-2 3.1.5 Cycle Operation ...........................
  • Page 8 Section Number Description Page Sterilizer Operation ........................6-1 Before Operating the Sterilizer ......................6-1 Preparing Loads for Sterilization Cycles ....................6-3 Guidelines for Placement of Various Loads................... 6-3 Loading & Unloading the Sterilizer ......................6-4 Loading Trolley and Carrier Instructions: Loading ................. 6-5 Loading Trolley and Carrier Instructions: Unloading................
  • Page 9 Section Number Description Page Routine Maintenance........................8-1 Preventive Maintenance Schedule ......................8-1 Daily Procedures ........................... 8-3 8.2.1 Clean Chamber Drain Strainer..................... 8-3 8.2.2 Clean Chamber ..........................8-4 Weekly Maintenance: Flush Chamber Drain ..................8-5 Maintenance As Required ........................8-5 8.4.1 Change Printer Paper Roll ......................8-5 8.4.2 Change Printer Ink Cartridge .......................
  • Page 10 Section Number Description Page 9.4.2 Too Long In Jacket Charge......................9-17 9.4.3 Too Long To Seal OE/NOE Door....................9-17 9.4.4 Too Long To Unseal OE/NOE Door................... 9-18 9.4.5 Chamber Pressure Transducer Failure..................9-18 9.4.6 Chamber Temperature Probe Failure ..................9-19 9.4.7 Jacket Temperature Probe Failure.....................
  • Page 11 LIST OF FIGURES Figure 5-1. Component Identification (660 x 660 mm Sterilizer Shown) ..............5-1 Figure 5-2. Sterilizer Key Switch and Countdown to Shutdown Timer..............5-3 Figure 5-3. Typical Out-of-Cycle Display ........................5-4 Figure 5-4. Typical In-Cycle and In-Cycle Status Display ..................5-4 Figure 5-5. Typical Alarm and Alarm Detail Screens ....................5-5 Figure 5-6.
  • Page 12 Table 7-2. Test Cycles ............................7-2 Table 7-3. Sterilizer Chamber Volume ........................7-2 Table 7-4. Change Machine Setup ........................7-14 ® Table 8-1. Preventive Maintenance Schedule for Amsco Evolution™ Medium Steam Sterilizers......8-1 Table 9-1. Operator Troubleshooting – In-Cycle Alarms ..................9-3 Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms................9-11 Table 9-3.
  • Page 13: Safety Precautions

    SAFETY PRECAUTIONS ® The following Safety Precautions must be observed when operating or servicing this Amsco Evolution™ Medium Steam Sterilizer. WARNING indicates the potential for personal injury and CAUTION indicates the potential for damage to equipment. For emphasis, certain Safety Precautions are repeated throughout the manual. It is important to review all Safety Precautions before operating or servicing the unit.
  • Page 14 Personnel must properly shutdown the sterilizer and retain the power key at all times while cleaning the sterilizer chamber. WARNING – SHOCK AND BURN HAZARD: Disconnect all utilities to sterilizer before servicing. Do not service the sterilizer unless all utilities have been properly locked out.
  • Page 15 CAUTION – POSSIBLE EQUIPMENT DAMAGE HAZARD: Allow thermostatic traps to cool down to room temperature before removing cover. Since there is nothing to limit expansion, the bellows may rupture or fatigue if trap is opened while hot. Actuation of safety relief valves at less than 75% of rated pressure can allow debris to contaminate the seat and cause the safety valve to leak.
  • Page 16 P129390-106 Operator Manual Safety Precautions...
  • Page 17: Symbol Definitions

    SYMBOL DEFINITIONS ® The following symbols appear on Amsco Evolution™ Series Medium Steam Sterilizers and are provided here for reference. Table 2-1. Symbol Definitions Symbol Definition Transfer of heat, hot surface Protective earth (i.e., ground) Electrostatic sensitive device Locked out cycle or machine set up option...
  • Page 18 Table 2-1. Symbol Definitions (Continued) Symbol Definition Standby Button Start Cycle Button Duplicate Print Button Abort Cycle Button Silence Audible Alarm Signal Button Display Abort Status Button Options Button Change Cycle Values Button OK (Accept) Button P129390-106 Operator Manual Symbol Definitions...
  • Page 19 Table 2-1. Symbol Definitions (Continued) Symbol Definition Backspace Button (erase entered characters) Exit Screen Button Machine Setup Button Adjust Audible Signals Button Set Audible Signal Off Set Audible Signal to Low Volume Set Audible Signal Medium Volume Set Audible Signal to High Volume Set Language Button Symbol Definitions Operator Manual...
  • Page 20 Table 2-1. Symbol Definitions (Continued) Symbol Definition Select English Language Button Select French Language Button Select German Language Button Select Italian Language Button Select Polish Language Button Select Spanish Language Button Select Finnish Language Button Set Utilities Control Button Lockout Button P129390-106 Operator Manual Symbol Definitions...
  • Page 21 Table 2-1. Symbol Definitions (Continued) Symbol Definition Set Print Options Button Set Machine ID Button Set Units Button Set Time and Date Button Set Date Button Set Time Button Display Chamber Temperature and Pressure Status Alarm Details Button Alarm Status Button Symbol Definitions Operator Manual P129390-106...
  • Page 22 P129390-106 Operator Manual Symbol Definitions...
  • Page 23: Installation Verification

    INSTALLATION VERIFICATION 3.1 Installation An Equipment Drawing showing all utility and space requirements was supplied with the sterilizer. Clearance space shown on the drawing is Checklist necessary for ease of installation and to assure proper operation and maintenance of equipment. Uncrating and Installation Instructions were also furnished with the sterilizer.
  • Page 24: Electrical Service

    ❑ The protective earth ground must be connected to terminal block 3.1.3 Electrical Service TB-1 in the sterilizer power box. ❑ Three-phase power for the sterilizer must meet specifications on the equipment drawing. ❑ Verify proper rotation of the vacuum pump by observing pump rotor shaft.
  • Page 25: Technical Specifications

    3.2 Technical Specifications 3.2.1 Overall Exterior Dimensions W x H x L Table 3-1. Sterilizer Overall Exterior Dimensions Sterilizer Model Exterior Dimensions (W x H x L) HC 600 Sliding Door 1321 x 2045 x 1626 mm (52 x 80-1/2 x 64") HC 800 Sliding Door 1321 x 2045 x 1880 mm (52 x 80-1/2 x 74")
  • Page 26: Utility Requirements

    3.2.3 Utility Requirements • Electric: 400 V, 3-phase, 6A / phase, 50 Hz • Water: Pressure: 1.4 to 3.5 bar (20 to 50 psig) Temperature: 15°C (59°F) maximum Consumption: 49 lpm (13 gpm), peak • Steam: Pressure: 3.5 to 5.5 bar (50 to 80 psig) Consumption, Peak: Approximately 152 kg/hr (335 lbs/hr) •...
  • Page 27: Techniques Of Sterilization

    TECHNIQUES OF STERILIZATION 4.1 General The information in this section is intended as a general guide to steam sterilization techniques. Also recommended is reference to the standards of the EN 285:2006, Sterilization – Steam Sterilizers, Large Steam Sterilizers EN ISO 17665-1:2006, Sterilization of Healthcare Products –...
  • Page 28: Control Measure For Verifying Sterilization Process

    Table 4-1. Cycle Information Sterilize Validation Cycle Type Load Sterilize Temp Time Time Standard Processing Cycles Sterilization Modules Prevac 134°C (273°F) 3 Min. 30 Sec. 20 Min. EN 285:2006 Table 4-2 (see below) Sterilization Modules Prevac 134°C (273°F) 3 Min. 30 Sec. 30 Min.
  • Page 29: Testing For Prevacuum Efficiency

    4.2.2 Testing for Prevacuum Tests such as the Bowie-Dick are designed to document the removal Efficiency of residual air from a sample challenge load. Run a Bowie-Dick test cycle daily before processing any loads in a WARNING – STERILITY sterilizer equipped with prevacuum cycles. The first prevacuum ASSURANCE HAZARD: Load cycle of each day should be used to test the adequacy of air removal sterility may be compromised...
  • Page 30: Vacuum Leak Test

    4.4 Vacuum Leak Test The Vacuum Leak Test (seeS 6.10.2, V ECTION ACUUM EST ON 6-22) measures the integrity of the sealed pressure vessel and PAGE associated piping to assure air is not being admitted to the sterilizer during the vacuum draw downs. After running a Leak Test cycle, a value or leak rate will be printed on WARNING –...
  • Page 31: Component Identification

    COMPONENT IDENTIFICATION Typical 660 x 660 mm (26 x 26”) Operator Interface Control Touch Screen Chamber Pressure Gauge Jacket Pressure Gauge Close (Raise) Door Button Open (Lower) Door Button Printer Emergency Stop Button Sterilizer Key Switch Control Panel Chamber Door (Closed) Location of Utility Supply Valves Piping Skid...
  • Page 32: General

    5.1 General Use this manual to become familiar with control locations and functions before operating the sterilizer (refer to Figure 5-1 through Figure 5-16). The controls for this sterilizer are contained within the control touch screen. Control touch pads appear on the screen as needed during each operation.
  • Page 33: Figure 5-2. Sterilizer Key Switch And Countdown To Shutdown Timer

    • Sterilizer Key Switch Refer to Figure 5-2. Key Switch Function: The key switch has two positions – 0 (Off) and 1 (On). When power is supplied to sterilizer, the key should be in the 0 (Off) position. When the key is moved to 1 (On) position, 24V dc power is applied to control and power-up sequence begins.
  • Page 34: Control

    5.3 Control Control displays can be divided into two categories, those occurring when the sterilizer is “out-of-cycle” and those occurring when the sterilizer is “in-cycle.” Typical out-of-cycle and in-cycle displays are shown in Figure 5-3 and Figure 5-4. • Out-of-cycle displays are used to start cycles, or set up and adjust sterilizer operation.
  • Page 35: Alarm Displays

    5.4 Alarm Displays Alarm displays tell operators and technicians when the sterilizer is experiencing an abnormal condition. Alarm conditions can be caused by failure of utility supplies or sterilizer components. 9, T , details the steps an operator can take ECTION ROUBLESHOOTING to solve most alarm conditions.
  • Page 36: Cycle Selection Touch Pads

    Cycle status and control messages are shown on a 213 mm (8-4/10 inch) color graphics display. Cycles can be started or aborted using the touch screen pads. Cycles and cycle values can be set using the Change Values procedure (accessible from the sterilizer cycle select screen).
  • Page 37: Change Values Control Screens

    ® The Amsco Evolution™ sterilizer control can be programmed to retain values for up to 12 separate cycles. The four cycles shown on the cycle select screen are the factory default cycles. Up to eight additional cycles can be programmed and displayed. It is the responsibility of the healthcare facility to validate any modification to cycles.
  • Page 38: System Shutdown Button

    5.6.2 System Shutdown The system shutdown button is used to turn off the sterilizer. Under Button normal circumstances the sterilizer should be left on at all times. Use the shutdown button to turn off the sterilizer when the machine must be serviced or for similar purposes.
  • Page 39: Abort Touch-Screen Button

    5.6.3 Abort Touch-Screen The Abort button is used to end a cycle before it completes normally. A cycle only needs to be aborted if an abnormal condition or a Button control problem develops during the cycle. Pressing Abort causes the sterilizer chamber to depressurize (if pressurized), or air break (if in vacuum);...
  • Page 40: Printer

    5.7 Printer Printer records all cycle data on 57 mm (2-1/4”) wide single-ply paper. See S 8.4.1. C 8-5 for ECTION HANGE RINTER APER PAGE paper changing procedure. Refer to Figure 5-14 for a summary of printer features. Take-Up Spool Inner Printer Door Paper Feed Button Printer Door...
  • Page 41: Printouts

    5.8 Printouts Refer to Figure 5-16. The printout reports useful information about each cycle the sterilizer runs; including the load number, which is a unique identifying code. Each load number is printed in the following format: a two-digit month (e.g., April = 04), a two-digit day (e.g., twenty eighth day = 28) and a two-digit daily cycle count (e.g., first cycle = 01, second = 02, etc.).
  • Page 42: Figure 5-16. Typical Printout

    Cycle Type Cycle Start Time & Date Total Cycle Count Operator I.D. Machine Number Serial Number Sterilize Temperature Control Overdrive Temperature Sterilize Time Dry Time Status Print Codes: Conditioning: • Charge • Vacuum/Pressure Pulses Sterilize Exhaust & Dry Complete Load Number Sterilize Temperature (Min./Max.) Time in Phase Total Cycle Time...
  • Page 43: Vertical-Sliding Door Operation (660 X 660 Mm Sterilizers)

    5.9 Vertical-Sliding The sterilizer door is operated using push buttons near the sterilizer control screen. Refer to Figure 5-1. Door Operation (660 x 660 mm • Pressing the "OPEN DOOR" push button while the door is in the Sterilizers) closed (up) position causes the door to open (by lowering). •...
  • Page 44 5-14 P129390-106 Operator Manual Component Identification...
  • Page 45: Sterilizer Operation

    STERILIZER OPERATION 6.1 Before Operating Operate sterilizer by referring to the cycle description in S ECTION 6.8, P 6-9. the Sterilizer REVAC YCLE PERATION PAGE 1. Press anywhere on the touch-screen of the sterilizer control WARNING – BURN HAZARD: display. Sterilizer, rack/shelves, and 2.
  • Page 46: Table 6-1. Prevacuum Sterilizer Cycle Values

    Table 6-1. Prevacuum Sterilizer Cycle Values Validation Cycle Sterilize Temp. Sterilize Time Dry Time Standard PREVAC 134°C (273°F) 3 Min., 30 Sec 30 Min. or 20 Min EN 285:2006 PREVAC 121°C (250°F) 30 Min. 30 Min. or 20 Min EN 285:2006 LIQUID 121°C (250°F) 45 Min.
  • Page 47: Preparing Loads For Sterilization Cycles

    6.2 Preparing Loads Before sterilization, all materials must be thoroughly cleaned. for Sterilization Cycles ® The Amsco Evolution™ sterilization chamber holds commonly used sterilization modules, wrapped or unwrapped instruments and equipment. 1. Wrappers may be made of 100% cotton (muslin), 140 thread count, two-ply fabric, and must be laundered;...
  • Page 48: Loading & Unloading The Sterilizer

    6.4 Loading & At the end of a cycle, when end-of-cycle tone sounds and display shows: Unloading the Sterilizer . . . open the chamber door. WARNING – BURN HAZARD: NOTE: Wear clean gloves and use clean towels as “pot holders” •...
  • Page 49: Figure 6-1. Loading Trolley And Carrier

    6.5 Loading Trolley 1. Open sterilizer door. and Carrier 2. Verify loading carrier is properly aligned upon loading trolley. Instructions: Loading 3. Align front end of loading trolley with dock at base of sterilizer. (See Figure 6-1 and Figure 6-2). 4.
  • Page 50: Figure 6-2. Loading Goods Into Sterilizer Chamber

    Loading Carrier Loading Trolley Trolley Docking Interface Handwheel Figure 6-2. Loading Goods into Sterilizer Chamber P129390-106 Operator Manual Sterilizer Operation...
  • Page 51: Loading Trolley And Carrier Instructions: Unloading

    6.6 Loading Trolley and Carrier Instructions: Unloading 1. Open chamber door. WARNING – BURN HAZARD: 2. Move loading trolley forward until it engages with dock. • Sterilizer, shelves and loading 3. Once loading trolley engages with dock, loading trolley latches carrier will be hot after cycle is run.
  • Page 52: Prevacuum Sterilizer Cycle

    6.7 Prevacuum Amsco Evolution Medium Steam Sterilizers are shipped with factory- set cycle values. The cycle sequence for these cycles are given in Sterilizer Cycle 6.8, P 6-9. Refer to Table 6-4, ECTION REVAC YCLE PERATION PAGE below, for factory set, qualified cycle settings.
  • Page 53: Prevac Cycle Operation

    6.8 Prevac Cycle The 134°C (273°F), or 121° (250°F), cycle is used for sterilizing double-wrapped instrument trays or fabric packs. Operation 1. Refer to S 6.1, B ECTION EFORE PERATING THE TERILIZER PAGE WARNING – BURN HAZARD: (at the beginning of this section) before running this cycle. •...
  • Page 54 NOTE: Countdown timer on the display is estimated; timer self- corrects estimated time at the beginning of each phase. VACUUM/PRESSURE PULSES #1 - #7 – Three vacuum pulses (drawing from 0.1 bar to 0.8 bar) are followed by four pressure pulses (0.1 bar to 1.1 bar).
  • Page 55 DRY – Start of dry is printed and display counts down dry time remaining. AIR BREAK – Chamber is returned to atmospheric pressure. RETRACT SEAL – A vacuum is drawn on the seal, retracting it from inner surface of door. COMPLETE –...
  • Page 56: Figure 6-3. Prevacuum Cycle Printouts (134°C [273°F] And 121°C [250°F])

    134°C (273°F) Prevacuum 121°C (250°F) Prevacuum Cycle Printout Cycle Printout Figure 6-3. Prevacuum Cycle Printouts (134°C [273°F] and 121°C [250°F]) 6-12 P129390-106 Operator Manual Sterilizer Operation...
  • Page 57 6-13 Sterilizer Operation Operator Manual P129390-106...
  • Page 58: Table 6-5. Factory-Set Liquid Cycle And Cycle Values

    Table 6-5. Factory-Set Liquid Cycle and Cycle Values The Amsco Evolution Medium Prevacuum Sterilizer is equipped with the following factory programmed liquid sterilization cycle and cycle values. Sterilize Validation Cycles Sterilize Temp. Time Time Recommended Load Standard 121°C (250°F) 45 Min.
  • Page 59: Liquid Cycle

    6.9 Liquid Cycle This cycle is used for sterilizing liquids in vented closures. 1. Refer to S 6.1, B ECTION EFORE PERATING THE TERILIZER PAGE WARNING – EXPLOSION before running this cycle. HAZARD: This sterilizer is not 2. See instructions for using the loading car/transfer carriage or designed to process flammable compounds.
  • Page 60: Figure 6-5. Typical Printout Of A Liquid Cycle

    Figure 6-5. Typical Printout of a Liquid Cycle 6-16 P129390-106 Operator Manual Sterilizer Operation...
  • Page 61: Test Cycles

    6.10 Test Cycles Test cycles are factory programmed on prevacuum sterilizers. These cycles are used to verify the sterilizer is functioning at optimum capability. Press the right arrow button at the lower right hand corner of the screen to advance to Test Cycles access screen (below). Enter the Access code (Default code is "0000";...
  • Page 62 Press the desired test cycle button (warm-up cycle, Bowie-Dick test, or leak test) to run the required cycle. 6-18 P129390-106 Operator Manual Sterilizer Operation...
  • Page 63: Bowie-Dick Test

    6.10.1 Bowie-Dick Test This cycle is used to conduct a Bowie-Dick test on sterilizers that use prevacuum cycles. This test is only applicable to sterilizers that use prevacuum cycles. The Bowie-Dick test is designed to document the removal of residual WARNING –...
  • Page 64 FAST EXHAUST – Start of exhaust is printed and chamber is exhausted to 0.28 bar (4 psig). DRY – Start of dry is printed and display counts down dry time remaining. AIR BREAK – Chamber is returned to atmospheric pressure. RETRACT SEAL –...
  • Page 65: Figure 6-6. Bowie-Dick Test Cycle Printout

    Figure 6-6. Bowie-Dick Test Cycle Printout 6-21 Sterilizer Operation Operator Manual P129390-106...
  • Page 66: Vacuum Leak Test

    6.10.2 Vacuum This cycle is used for testing vacuum integrity of sterilizer and piping. Leak Test A Vacuum Leak Test cycle should be run on the sterilizer at least WARNING – BURN HAZARD: once each week. It should be one of the first cycles run for the day. •...
  • Page 67: Figure 6-7. Leak Test Cycle Printout

    Figure 6-7. Leak Test Cycle Printout 6-23 Sterilizer Operation Operator Manual P129390-106...
  • Page 68: Aborting Cycles

    6.11 Aborting Cycles It may be necessary to end a processing cycle, possibly because the wrong cycle was selected or the sterilizer begins functioning incorrectly. A cycle can be aborted at any time by pressing the Abort touch-screen button. 1. Touch the Abort touch-screen button. WARNING –...
  • Page 69 2. Once chamber reaches 0.28 Pbar (4 psig), the sterilizer removes vapor from the chamber for one minute. Press the silence alarm button to turn off the alarm speaker during the abort sequence. Press the status touch pad in the lower left hand corner of the screen at any time to display an abort- status screen (typical sample below): 6-25...
  • Page 70: Cycle Graph

    S 6.4, L & U ECTION OADING NLOADING 6-4. TERILIZER PAGE 6.12 Cycle Graph Prevacuum cycle graph for the Amsco Evolution Medium Steam Sterilizer is shown in Figure 6-8 (page 6-27). 6-26 P129390-106 Operator Manual Sterilizer Operation...
  • Page 71: Figure 6-8. Cycle Graph: Prevacuum Cycle - 134°C (273°F) And 121°C (250°F)

    Figure 6-8. Cycle Graph: Prevacuum Cycle – 134°C (273°F) and 121°C (250°F) 6-27 Sterilizer Operation Operator Manual P129390-106...
  • Page 72: Figure 6-9. Cycle Graph: Liquid Cycle

    Figure 6-9. Cycle Graph: Liquid Cycle 6-28 P129390-106 Operator Manual Sterilizer Operation...
  • Page 73: Cycle And Control Value Programming

    CYCLE AND CONTROL VALUE PROGRAMMING Amsco Evolution™ sterilizers are shipped with factory-set cycles, ® cycle values and control values programmed into the control (see Table 7-1). These preset values can be changed to tailor the sterilizer to the operating environment in which it has been installed.
  • Page 74: Cycle Values

    7.1 Cycle Values Amsco Evolution sterilizers are shipped with the default cycle values shown in Table 7-1. The Prevacuum Configuration Amsco Evolution sterilizer is equipped with the following factory programmed sterilization cycle values (Table 7-1). Table 7-1. Prevacuum Sterilizer Cycle Values...
  • Page 75: Change Values

    7.2 Change Values Important: Applicable cycles have been validated to satisfy the requirements outlined in Table 4-1 (page 4-2). If cycle parameters (sterilize time or dry time) other than those in Table 7-1 are required, it is the responsibility of the healthcare facility to validate the cycle. Reference EN ISO 17665-1:2006 for a guide to validating sterilization cycles and to ensure drying efficiencies are met.
  • Page 76: Change Cycle Values

    7.3 Change Cycle Important: Applicable cycles have been validated to satisfy the requirements outlined in Table 4-1. If cycle parameters (sterilize time Values or dry time) other than those in Table 7-1 are required, it is the responsibility of the healthcare facility to validate the cycle. Reference EN ISO 17665-1:2006 standards for a guide to validating sterilization cycles and to ensure drying efficiencies are met.
  • Page 77: Change Cycle Values Procedure

    7.3.2 Change Cycle Values Procedure Press anywhere on standby screen to access the sterilizer’s control functions. Touch Anywhere to Start Options Button Press the options button to access screen used to adjust sterilizer’s operation. Press change cycle values to adjust Change Cycle Values Button the operating parameters for specific...
  • Page 78 Continued from Previous Page Press value to be changed (Cycle Type, Sterilize Temperature, Sterilize Time or Dry Time). Once a value has been accepted at any individual change values screen (see below), pressing OK returns control to this screen. Press (backspace) to clear values that have been entered. Press (OK) to accept values that have been entered, and return to change cycle values screen.
  • Page 79: Change Time And Date

    7.4 Change Time and These screens adjust the time and date the sterilizer uses for all display and printout messages. Date The current time and date appears on the Standby and Cycle Status screens. Time and date are also shown on printouts. These should be verified periodically.
  • Page 80: Set Time

    changes to show the Options screen. Press Set Time & Date touch pad, the display advances to Set Time & Date screen. 2. Press Set Time Format touch pad. The format shown on the touch pad toggles through two available options: •...
  • Page 81 a. If an incorrect number is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. b. Once the correct time has been entered, press OK to accept change hours screen. c. Press EXIT to return to Change Time and Date screen. 4.
  • Page 82 b. Once the correct minutes value has been entered, press OK to accept. c. Press EXIT to return to Change Time and Date screen. 5. Change Seconds – use the touch screen key pad to enter the seconds value, this can be between 00 and 59. a.
  • Page 83: Date Format

    7.4.3 Date Format This setup option allows the operator to select the “format” for the date. The format determines the order in which the month, day and year are displayed. There are three options, and the option selected is a matter of either preference or geographical location. 1.
  • Page 84: Set Date

    7.4.4 Set Date At Change Time and Date screen, press Set Date and screen shown below appears. Enter the day value using the keypad. The day value must be between 1 and 31. 1. If an incorrect number is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value.
  • Page 85 Enter the correct month value using the keypad. The month value must be between 1 and 12. 1. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. 2.
  • Page 86: Change Machine Setup

    1. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. 2. Once the correct year value has been entered, press OK to accept. NOTE: The day of the week is automatically understood and registered by the control.
  • Page 87: Utility Control

    7.5.1 Utility Control The Utility Control screen is used to automatically control utility services to the sterilizer. As shipped from the factory, this is set to Manual Utilities control (i.e., utilities must be shut off and turned on by an operator). If the sterilizer is equipped with an integral electric steam generator, refer to S 7.5.3, U ECTION...
  • Page 88 The setting on this screen is either Manual or Automatic. If the utility control button displays Manual (and Automatic operation is preferred), press the utility control button to change setting to Automatic. If Automatic is displayed on the button (and Manual operation is preferred), press the button to change to Manual.
  • Page 89 1. Press the Day of the Week button to select the day to set up for utility control. • The Day of the Week button sequentially displays the days of the week each time it is pressed. • Press the button repeatedly until reaching the day of the week to be set up.
  • Page 90 4. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. 5. Press Exit to abandon changes and return to the Machine Setup screen. Press OK to save utilities set up changes and return to Utility Control screen.
  • Page 91 8. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value. 9. Press Exit to abandon changes and return to the Machine Setup screen. Press OK to save utilities set up changes and return to Utility Control screen.
  • Page 92: Utility Control Summary

    7.5.2 Utility Control Summary The utility control function allows the operator to display a summary of all the utility controls settings. 1. To display a summary of the automatic utility control restart and shutdown times that have been entered, press the information key in the upper left-hand corner of the Utility Control screen.
  • Page 93: Utility Control With Integral Electric Steam Generator

    7.5.3 Utility Control with If the sterilizer is set up to be equipped with an integral electric Integral Electric Steam steam generator, the following screen appears on the display when Utility Control is accessed. This screen adds the option for setting a Generator daily generator flush.
  • Page 94: Language

    7.5.4 Language The Amsco Evolution sterilizer is capable of operation with display screens and printouts in seven languages. The factory default is English. 1. To access this utility, press the options button from the main status screen at the operating end of the sterilizer. The screen changes to show the Machine Setup screen.
  • Page 95 3. Select the appropriate language by pressing one of the buttons in the middle of the display. Language buttons are represented by the country flag where the language is spoken. The buttons and corresponding languages are listed below. English Language French Language German Language Italian Language...
  • Page 96: Machine Number

    7.5.5 Machine Number This is used to enter an identifying, two-character code into the sterilizer control. This code can be any two-digit number between 00 and 99. The Machine Number code is printed out in the header for each cycle printout, allowing for processed goods to be traced back to a specific sterilizer when needed.
  • Page 97 3. At machine number screen, enter a two-digit code for the sterilizer using the entry keypad. Ensure, however, that each machine number used is different from any others that have been assigned in the facility. 4. If an incorrect value is entered, press the backspace key to clear the entry, then use the keypad to enter the correct value.
  • Page 98: Print Options

    7.5.6 Print Options This setup option allows the operator to select the cycle printout “format.” The format determines the type of printout the sterilizer provides during processing. Two options are available. The default Full format provides status prints at each transition point in the cycle, plus additional status at interval points during each phase of the cycle.
  • Page 99 1. Select the appropriate print format by pressing one of the three buttons on display. • Print Format – Use this option to set the type of print format provided by the sterilizer during cycle operation: • Full – This is the standard format providing a status print for each phase of the cycle and status prints at a predetermined print interval.
  • Page 100: Audible Signals

    7.5.7 Audible Signals This setup option allows the operator to adjust selected audible signals heard at the sterilizer control. Three signals can be adjusted. Touch pad and end of cycle signals can be adjusted to one of three volume levels or turned off. Only the volume level of the Alarm signal can be adjusted.
  • Page 101 3. Select the signal you wish to adjust by pressing the appropriate touch-screen pad in the upper half of the screen. Press each button multiple times to cycle through the available volume levels (including Off, when applicable) for the selected signal type: NOTE: TOUCHPAD and END OF CYCLE provide an OFF setting.
  • Page 102: Units

    This function allows selection of either Fahrenheit or Celsius units for displaying and printing temperature. Pressure units can be changed between psig/inHg or bars. NOTE: Although Amsco Evolution Medium Steam Sterilizers are factory-set to operate using Celsius temperature units, the sterilizer can be reprogrammed to display and print using Fahrenheit temperature units.
  • Page 103: Leaving Machine Setup

    3. At units screen, press the appropriate button for the type of unit or units to be selected. • Press Temp Units button to toggle between Celsius (°C) and Fahrenheit (°F). • Press Pressure Units button to toggle between PSIG and bar.
  • Page 104 7-32 P129390-106 Operator Manual Cycle and Control Value Programming...
  • Page 105: Routine Maintenance

    Contact your STERIS Service Representative to schedule preventive maintenance. ® Table 8-1. Preventive Maintenance Schedule for Amsco Evolution™ Medium Steam Sterilizers Service Required Minimum Frequency PREPARATION FOR PREVENTIVE MAINTENANCE Discuss equipment with operators and check printouts.
  • Page 106 ® Table 8-1. Preventive Maintenance Schedule for Amsco Evolution™ Medium Steam Sterilizers (Continued) Service Required Minimum Frequency Inspect all check valves. Repair/replace as needed. 1x per year Rebuild steam control valve (PRV). 1x per year Verify that safety valve is not leaking.
  • Page 107: Daily Procedures

    ® Table 8-1. Preventive Maintenance Schedule for Amsco Evolution™ Medium Steam Sterilizers (Continued) Service Required Minimum Frequency 5.11 Verify pressure settings as described in Maintenance Manual. 1x per year SAFETY TESTING Inspect ground bond. 1x per year Inspect steam connection to sterilizer.
  • Page 108: Clean Chamber

    8.2.2 Clean Chamber Important: • Chamber must be at room temperature, sterilizer off all night, WARNING – SLIPPING before washing. HAZARD: To reduce slippery • The entire chamber should be wiped down and rinsed floor conditions, keep floors following any spills or other soiling. dry by immediately wiping up any spilled liquids or 1.
  • Page 109: Weekly Maintenance: Flush Chamber Drain

    8.3 Weekly Flush chamber drain as follows whenever the line becomes clogged: Maintenance: 1. Turn off steam supply valve. Wait until jacket pressure is zero. Flush Chamber Drain Wait until chamber has cooled to room temperature. 2. Remove chamber drain strainer. Clean strainer using WARNING –...
  • Page 110: Figure 8-2. Printer Components (Printer Door Open)

    Take-Up Spool Inner Printer Door Paper Feed Button Printer Door Figure 8-2. Printer Components (Printer Door Open) 2. Tear paper between take-up spool and printer. Refer to Figure 8-3 Figure 8-3. Tear Printer Paper 3. Remove take-up spool from drive by inserting fingers in cavity and lifting slightly to pull spool out.
  • Page 111: Figure 8-4. Insert New Paper Roll

    6. Insert new paper roll. Refer to Figure 8-4. Figure 8-4. Insert New Paper Roll 7. Insert end of paper into printer roller mechanism just behind ink cartridge. Feed Paper Into Roller Mechanism Figure 8-5. Insert Paper Behind Printer Ink Cartridge Routine Maintenance Operator Manual P129390-106...
  • Page 112: Figure 8-6. Press Paper Freed Button To Advance Paper

    8. Press paper feed button on printer until paper advances through printer mechanism and ink cartridge, exiting through front. Advance Paper by Pressing Paper Feed Button Figure 8-6. Press Paper Freed Button to Advance Paper 9. Continue pressing paper feed button (or pull paper gently) until about 46cm (18”) of paper hangs out of printer.
  • Page 113: Change Printer Ink Cartridge

    8.4.2 Change Printer Ink The printer ink cartridge should be changed as soon as type on Cartridge printouts is light or faded, and before printouts become difficult to read. 1. Tear paper between take-up spool and printer. 2. Open access door, then press on right end of ink cartridge, until left end of cartridge pops out of the printer.
  • Page 114 5. Retighten ribbon by rotating wheel on left side of cartridge 1/4 8.4.1, C turn. Then see S ECTION HANGE RINTER APER 8-5, Step 9., through Step 12., to reinstall take-up spool. 8-10 P129390-106 Operator Manual Routine Maintenance...
  • Page 115: Troubleshooting

    TROUBLESHOOTING 9.1 General This section pictorially lists and describes alarm conditions that may ® occur when operating the Amsco Evolution™ Medium Steam Sterilizer. NOTE: Never permit unqualified persons to service the sterilizer. WARNING – PERSONAL Important: The Service Information provided in the following pages...
  • Page 116: Typical Alarm Screen

    9.1.1 Typical Alarm Screen When an alarm condition occurs, the alarm tone sounds and the touch screen automatically displays the corresponding alarm screen. Typically, each alarm screen indicates the alarm name, the time of day and the date. The alarm details screen (if available), provides more information, including operator instructions.
  • Page 117: Typical Alarm Printout

    9.1.2 Typical Alarm Printout When an alarm occurs the printer automatically generates a printout, typically listing alarm name, time alarm occurred, current chamber status, and any associated sensor temperature. See Figure 9-2. * Alarm XX/XX/XX PRESSURE IN CHAMBER F XX:XX:XXA/P X:XX A/P Full Print Format Shown Figure 9-2.
  • Page 118: Too Long In Exhaust

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.2 Too Long In Alarm Screen Chamber did not exhaust to atmospheric pressure within Exhaust allotted time. Occurs if chamber does not Causes and Correction: exhaust to 0.28 bar (4 psig) 1.
  • Page 119: Too Long In Evacuation

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.3 Too Long In Alarm Screen Chamber reach Evacuation required vacuum level within allotted time. Occurs if chamber does not Causes and Correction: reach the set evacuation 1.
  • Page 120: Too Long In Air Break

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.4 Too Long In Alarm Screen Chamber did not air break vacuum to 0.1 vbar (2 inHg) Air Break within allotted time Occurs if chamber does not Causes and Correction: air break the vacuum to 1.
  • Page 121: Over Sterilize Temperature

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.6 Over Sterilize Alarm Screen Sterilize temperature is above setpoint more than Temperature prescribed amount. Occurs if chamber Causes And Correction: temperature exceeds the 1.
  • Page 122: Chamber Pressure/Temperature Failure

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.8 Chamber Alarm Screen Pressure temperature outside normal steam range. Pressure/ Temperature Failure Causes and Correction: 1. Control out of calibration Occurs if chamber pressure •...
  • Page 123: Steam Utility Failure

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.10 Steam Utility Alarm Screen Check steam utility supply restore proper Failure operation. Steam supply to sterilizer is Causes and Corrections: shut off or has failed. 1.
  • Page 124: Air Utility Failure

    Table 9-1. Operator Troubleshooting – In-Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.2.12 Air Utility Alarm Screen Check air utility supply and restore to proper operation. Failure Causes and Corrections: Air supply to sterilizer is 1. Air pressure less than shut off or has failed.
  • Page 125: Out-Of-Cycle Alarms

    9.3 Out-of-cycle The following alarm screens will appear only when the sterilizer is not processing a cycle. Alarms Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms Condition Alarm Screen on Sterilizer Control Service Information 9.3.1 Too Long To Alarm Screen Door switch did not make contact in allotted time.
  • Page 126 Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.3.2 Too Long To Alarm Screen Door switch did not open in allotted time. Open OE (or NOE) Door Causes and Correction: 1. Door switch malfunction Occurs if door switch does •...
  • Page 127: Rtc Failure

    Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.3.4 Waste Alarm Screen RTD probe, RTD4, output is outside normal range. Temperature Probe Failure Causes and Correction: 1. Loose connections in probe Occurs if waste line wiring temperature reading is •...
  • Page 128 Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.3.6 Emergency Alarm Screen The emergency stop button was pressed either in or out of cycle. Stop Button Pressed Causes and Correction: 1. Correct the emergency The emergency stop condition (if any) that button has been pressed,...
  • Page 129: Ups Battery Low

    Table 9-2. Operator Troubleshooting – Out-of--Cycle Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.3.8 UPS Battery Alarm Screen UPS battery or UPS may be malfunctioning. Causes and Correction: Battery power for 1. Battery failure uninterruptibile power • Replace supply has dropped below a level capable of...
  • Page 130: Sensor Alarms

    9.4 Sensor Alarms The following alarm screens appear anytime the sterilizer is energized. The sensors are continually monitored whenever sterilizer is in- or out-of-cycle. Table 9-3. Operator Troubleshooting – Sensor Alarms Condition Alarm Screen on Sterilizer Control Service Information 9.4.1 Water In Alarm Screen Excess water sensed in chamber, Chamber...
  • Page 131 Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.2 Too Long In Alarm Screen Jacket did not reach required temperature within allotted time. Jacket Charge Causes and Correction: Occurs if jacket does not 1.
  • Page 132 Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.4 Too Long To Alarm Screen Door seal pressure not below 0.4 bar (5 psig) within allotted Unseal OE/NOE time. Door Causes and Correction: Occurs door seal...
  • Page 133 Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.6 Chamber Alarm Screen RTD probe, RTD1, output is outside normal range. Temperature Probe Failure Causes and Correction: 1. Loose connection in probe Occurs if chamber wiring temperature reading •...
  • Page 134 Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.8 OE/NOE Door Alarm Screen Door switch open while seal switch closed. Switch Failure Causes and Correction: Occurs if OE or NOE door 1. Door switch malfunction seal switch contact is •...
  • Page 135: Too Long To Seal Oe/Noe Door

    Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.10 OE/NOE Alarm Screen Door switches do not agree. Door Switches Causes and Correction: Malfunction 1. Door switch malfunction • Check LS1(LS3) connections Occurs if control senses for OE door or LS2 (LS4) For OE or NOE door is in the NOE Door...
  • Page 136: Exhaust Rate Too Fast

    Table 9-3. Operator Troubleshooting – Sensor Alarms (Continued) Condition Alarm Screen on Sterilizer Control Service Information 9.4.12 Exhaust Alarm Screen Chamber exhausted faster than the expected rate. Rate Too Fast Causes and Correction: Occurs if liquid cycle fast 1. Valve malfunction – Repair exhaust rate is too fast.
  • Page 137: Service Procedures

    SERVICE PROCEDURES 10.1 General The material in this section is provided to allow for servicing WARNING – PERSONAL components of the sterilizer most likely to need attention. These INJURY HAZARD AND/OR procedures are more advanced than cleaning and replacing EQUIPMENT DAMAGE expendables (such as printer paper and door seals).
  • Page 138: Clean Jacket Steam Strainers

    10.3 Clean Jacket The strainers should be opened for cleaning after initial start-up and at least twice a year thereafter (refer to Table 8-1. Preventive Steam Strainers Maintenance Schedule). Accumulation of sediment and rust will reduce pressure and flow. In extreme conditions, complete blockage WARNING –...
  • Page 139 NOTE: Reference indicators are located inside the rear groove of door seal, at the middle of each side (refer to Figure 10-1 for 660 x 660 mm [26 x 26”] sterilizers, or Figure 10-2 for 660 x 950 mm [26 x 37.5”] sterilizers). b.
  • Page 140: Figure 10-1. Door Seal Replacement, 660 X 660 Mm (26 X 26") Sterilizers

    Chamber Opening Reference Indicator, Enlarged Location of Seal Reference Indicators Seal The following lot data is molded into the back side of Back Side the seal: of Door Seal P-129373-376 ABCDEFHGIJKLMNOP 1234 Part Number Lot Index Cure Year Manufacturer Quarter Code The lot index and cure quarter are struck off during manufacturing.
  • Page 141: Figure 10-2. Door Seal Replacement, 660 X 950 Mm (26 X 37.5") Sterilizers

    Chamber Opening Door Use flat tool with rounded edges (such as a non- serrated table knife) to pry and twist one section of the seal partially from the groove. Reference Indicator, Enlarged Seal Hinged Door Model Shown Location of Seal Reference Indicators The following lot data is molded into the back side of...
  • Page 142: Steam Trap Replacement

    10.5 Steam Trap Replacement Refer to Figure 10-3. WARNING – BURN HAZARD: Disassembly • Allow sterilizer and 1. Using a suitable wrench, unscrew and remove the cap and accessories to cool to room temperature before performing bellows assembly. any cleaning or maintenance 2.
  • Page 143: Safety Valve Test

    10.7 Safety Valve Test The safety valves are to be tested periodically (refer to Table 8-1. Preventive Maintenance Schedule at the beginning of S ECTION OUTINE AINTENANCE • Prevent damage during testing by ensuring that at least 75% of WARNING – BURN HAZARD: the rated pressure is in the chamber.
  • Page 144 Table 10-1. Recommended Spare Parts 093911351 ....RTD, Jacket Temp; 093921381 ....RTD, Waste Water Temp 093922107 ....RTD, Dual Chamber Temp; Recorder Temp 093927069 ....SWITCH, Chamber Flooded 093931004 ....OUTPUT MODULE, 24V dc, 8 Point dc 093931005 ....RTD, Input Module, 2C 093931013 ....CIRCUIT BREAKER, 2 Pole, 10a 093931017 ....EXPANSION POWER SUPPLY 093931020 ....CURRENT INPUT MODULE 093931021 ....MODULE, Relay Output...
  • Page 145: Figure 10-4. Evolution Sterilizer Piping Schematic

    Figure 10-4. Evolution Sterilizer Piping Schematic 10-9 Service Procedures Operator Manual P129390-106...
  • Page 146: Waste Products Disposal

    10.9 Waste Products The following are waste materials associated with the sterilizer. When disposing of waste materials, be sure to do so in compliance with Disposal federal, state, and local regulations. • Printer paper – recyclable. • Printer ribbon – not recyclable. •...