96 Deep-Well Plate to enhance the lysis and elution efficiency. Applications ZiXpress 64 DX system purifies DNA and RNA from a variety of samples which are essential and broadly used in molecular biology, genetic screening, sequencing, food safety, forensic, etc.
For your safety and those of others, follow the guidelines provided in the following pages concerning the use of the ZiXpress DX64 system. ZINEXTS LIFE SCIENCE CORP. will not be held liable for any errors, damage, or other unexpected events resulting from miscarried following safety guideline.
◼ system. Also, refrain from bumping or knocking it. ◼ Repairs to the ZiXpress 64 DX system should be only performed by representatives / agents that are specifically authorized, and with original spare parts that are certified by the ZINEXTS LIFE SCIENCE CORP.
Page 6
Note: ◼ ◼ Reagent kits will be generally not supplied with the ZiXpress 64 DX system instrument. Please select your desired Reagent Kits and order them separately. About Samples Biohazard: Always wear appropriate gloves, a mask, and safety goggles etc.
This product and any of its accessories is for the extended use of nucleic acid extraction, purification isolation in molecular biology research laboratories and clinical diagnosis laboratories. Specifications Product ZiXpress 64 DX system Model Name Specifications Bench-top automated Nucleic Instrument Type Acid Extractor...
INSTALLATION Composition of the ZiXpress 64 DX system Packaging Check that the following parts are included in the package. Contact your local representative if you notice any missing part(s). Instrument Quantity: Instrument x1 B. User Manual C. Power Cord x1 D. 2.2ml Deepwell 96 Plate x4 8-Tip Comb x4 F.
Page 12
transportation damages or improper handling and operation. Extraction Kits Quantity: 2.2ml Deepwell 96 Plate x12 (plates) 8-Tip Comb x24 (pcs) Note: Extraction kits have to be purchased separately. Please contact your local representative / agent for further information. The contents of extraction kits are various. Refer to the Reagent Kit Handbook, enclosed in the reagent box, for further details.
Operating Environment / Conditions ZiXpress 64 DX Use the ZiXpress 64 DX system system in an environment system that meets the following conditions: Operating Environment ◼ The space for the allocation of ZiXpress 64 DX system must secure 5-10 cm distance from the instrument to the wall.
Page 14
Caution: Do not use the ZiXpress 64 DX system in an unstable place such as a slanted surface or a place subject to vibrations. It may cause injury or device failure.
Settings Important: The ZiXpress 64 DX system has a weight of more than 25kgs. It should be lift and moved by two persons. Hold the moving handler of the instrument from two sides to move it out of the transportation box.
Page 16
Keep the shipping box and fixing block, they are needed for transporting the ZiXpress 64 DX system. (2) Connect power cord to the instrument. Power inlet (3) Remove the Fixing Block inside the instrument and place it in the Fixing Block Sensor Spot, following the guideline below:...
Notes for Transportation / Shipping ZiXpress 64 DX When transporting / shipping the ZiXpress 64 DX system to system a new location, perform the following tasks: Transportation 1. Insert Fixing Block (1) Put the Fixing Block to fix the Magnetic Pillar frame.
3. Pack the ZiXpress 64 DX system instrument, including accessories in its original shipping box or some other equivalent parcel. Functional Descriptions Main Menu Home Light Light Setting Hibernate LED Light After pressing button, the icon will change into the instrument’s light will turn off.
Page 20
Settings After pressing button, you can do the instrument adjustment / check / test. After pressing button, the instrument will get into the Sleeping mode “sleeping mode’’. If you want to wake up the instrument, touch the screen again. Note: Before getting into the sleeping mode, please press button first to turn off the LED light.
GETTING STARTED Preparation Biohazard: Always wear appropriate gloves, a mask, and safety goggles during any biohazardous operations in extraction process. Even when touching the device after any operation with a biohazard risk, wear appropriate gloves and a mask since the device may be contaminated.
(2) Connect the plug of the power cable to the outlet. (3) Turn the power switch on and wait for the Touch Panel to turn on and displaying “ZiXpress 64 DX system initializing”. The system will process self-testing, and then go to steady mode.
Page 23
(4) Open the door. (5) Set up the mixing sleeves on the mixing sleeve track. Caution: Make sure the sleeves have entered completely! Page 22...
Page 24
(6) Remove carefully the aluminum foil sealing membrane on the reagent plate. Add samples to Well 1 and Well 7. (7) Place the reagent plate(s) into the instrument. Page 23...
Page 25
(8) Close the door before starting the instrument! (9) Select protocol Select the pre-installed protocols. This figure is the standard protocol in ZiXpress 64 DX system: Page 24...
Page 26
(10) Press to start the extraction process. Note: The process time and remaining process time will be displayed on the screen. (11) While the extraction procedure is finished, transfer the extracted products located at Well 6 and Well 12 into nuclease-free tubes.
Page 27
Note: Store the purified nucleic acid at 4°C (short-term, less than 10 days) or aliquot and store at -70°C (long-term) before performing any downstream analysis. (12) Discard the used consumables into biohazard waste. (13) If you are not using the instrument in short-term, press button to get into “sleeping mode”.
Page 28
(2) To restart the procedure, press button Stop the Process (1) During the process, if you want to stop/pause the procedure, please press button. (2) To stop the process, press button, Page 27...
Page 29
(3) Press <Yes>, then the instrument will do the initializing. After Extraction (1) Transfer the extracted products located at Well 6 and Process Well 12 into nuclease-free tubes. Note: You can apply quality checking, do downstream study or storage them as what you expect. (2) Remove the used plastic disposables from the instrument and dispose them.
If the outside of the ZiXpress 64 DX system is heavily soiled, or if any samples that may cause infection are adhered to the outside of the ZiXpress 64 DX system, wipe with a soft paper tissue, soaked with 75% ethanol. ZiXpress 64 DX...
TROUBLESHOOTING Problem Cause Solution Check AC power cord Power disconnected connections at both ends. Or use the correct cords. No display after Blown fuse Replace the fuse turned on the power switch. Malfunction power switch Replace the switch Contact your local Others representative / agent Touch Panel turns...
Mail: info@zinexts.com Product of Origin: Taiwan (R.O.C.) Technical Support Zinexts Life Science - Service Center For technical problem and instrument maintenance please contact our service center: support@zinexts.com Before contacting the Technical Support, refer to “Troubleshooting” to check the problems. If the issue cannot be solved, contact your local representative / agent, or the Zinexts Life Science - Service Center.
(ii) Any consumption spare parts including fuses, light bulbs, UV lights, LED lighting sets. For further information about the Warranty Policy and Extended Warranty offers, please contact your local representative / agent, or visit our website www.zinexts.com. Page 32...
Page 34
Zinexts Life Science Corp. www.zinexts.com 16F., No. 93, Sec. 1, Xintai 5th Rd., Xizhi Dist., New Taipei City 221416 , Taiwan (R.O.C) Tel: +886 2 2246 3579 Fax: +886 2 2243 8570 Mail: info@zinexts.com Page 33...