ZINEXTS MagPurix 12 EVO Instruction Manual

ZINEXTS MagPurix 12 EVO Instruction Manual

Automated nucleic acid purification system

Advertisement

Quick Links

Automated Nucleic Acid Purification System
Catalog No.: ZP01018
Rev. Date: Sep. 07 2022
Manual No. IFU-MP01-01018
Version: 3.3
Obelis s.a.
Bd Général Wahis 53
1030 Brussels Belgium
Tel: +(32) 2 732-59-54
Fax: +(32)2 732-60-03
mail@obelis.net
Instruction Manual
12 EVO

Advertisement

Table of Contents
loading

Summary of Contents for ZINEXTS MagPurix 12 EVO

  • Page 1 Instruction Manual Automated Nucleic Acid Purification System 12 EVO Catalog No.: ZP01018 Rev. Date: Sep. 07 2022 Manual No. IFU-MP01-01018 Version: 3.3 Obelis s.a. Bd Général Wahis 53 1030 Brussels Belgium Tel: +(32) 2 732-59-54 Fax: +(32)2 732-60-03 mail@obelis.net...
  • Page 2: Table Of Contents

    Introduction ............................... 3 Safety Information ........................... 4 1.0 Installation ............................8 1.1 Composition of the MagPurix 12 EVO system ................ 8 1.2 Contents of Reagent Kits ......................9 1.3 Operating Environment / Condition ..................9 1.4 Unpack the MagPurix 12 EVO system ..................11 1.5 Overall View ..........................
  • Page 3: Introduction

    Introduction Thank you for purchasing the MagPurix 12 EVO which is a fully automated, standalone robot that can purify nucleic acids within 30-45 minutes. With advanced magnetic bead separation technology, it enables you to obtain high-quality extraction results. Moreover, the most user-friendly interface makes users free from troublesome parameter settings and maintenance.
  • Page 4: Safety Information

    This symbol shows this instrument is an “In vitro diagnostic”-certified medical device. For your safety and that of others, follow the guidelines provided in the following pages concerning the use of the MagPurix 12 EVO system. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 5 Do not use the MagPurix 12 EVO system with a damaged power plug or a loose socket. If there is dust on the prongs of the power plug or on the plug socket, remove it with a dry cloth.
  • Page 6 Extraction should be performed in an appropriate laboratory or workplace.  Note: The kits are not supplied with the MagPurix 12 EVO system. Select the desired kit(s) and  order it (them) separately. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 7 Zinexts Life Science Corp. www.zinexts.com 16F., No. 93, Sec. 1, Xintai 5th Rd., Xizhi Dist., New Taipei City 221416, Taiwan Tel: +886 2 2246 3579 Fax: +886 2 2243 8570 www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 8: Installation

    Please make sure all the components are free from damages as soon as you get the system. If any damage is found, please contact your local representative for instant support. Note: Zinexts’ global warranty does not cover damages from transportation or improper  operation. www.zinexts.com...
  • Page 9: Contents Of Reagent Kits

    The contents of reagent kits will vary. Refer to the handbook of kits enclosed in  reagent box for details. 1.3 Operating Environment / Condition Use the MagPurix 12 EVO system in a location that meets the following conditions: www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 10 The MagPurix 12 EVO system must be place at a minimum distance of 5-10 cm  from all sidewalls to allow good air circulation. The space for the allocation of the MagPurix 12 EVO system must remain 5-10 cm from instrument to sidewalls.
  • Page 11: Unpack The Magpurix 12 Evo System

    Warning: Do not use the MagPurix 12 EVO system in a location where it is wet or can be splashed with water. It may cause device failure, fire, or electric shock. When relocating the MagPurix 12 EVO system, disconnect the plug from the outlet first.
  • Page 12: Overall View

    Improper MagPurix 12 EVO system movement handling will lead to instrument damages. For correct and safe use of the MagPurix 12 EVO, install the device in a location close to the electrical outlet and that has enough space for installation and main switch operation.
  • Page 13 UV Lamp Camera Operation Area There is a total of 12 strips (from left to right) Inside view  Piston Reagent kit holder Heating module Tip holder plate M module Holder lid Sample rack www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 14: Getting Started

    “Service” password to lock access to end user to certain “service” options) (Please refer to System Status for the operate information) iii. Set up the local time. (Please refer to System Status for operating information) www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 15 Dashcam time adjustment is suggested for the first use. Fallow the steps below to set the dashcam time: Step 1, remove the dashcam as below. Step 2, turn on instrument and find System from Settings. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 16 Step 6, long press “M” button access to recording after turn off sensor. Press menu. “OK “button to pause. Step 7, press “M” button again to move to Step 8, select Date/Time and press OK the next page. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 17 OK button to return to the main screen. Step 11, back to panel and click camera to open sensor. And then hang camera back follow the same way. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 18: Touch Panel And User Interface

    Once logged in, you will able to set up at least 10 other users with their own password. Important: Keep the shipping box and fixing block materials. They will be needed again when transporting the MagPurix 12 EVO system. 2.2 Touch Panel and User Interface Function Bar Homepage Status Bar 2.3 Preparation...
  • Page 19: Function Test

    Enter the settings menu and check the position of each axis (Y/Z/P/M). Press button first then axis will go back to home position Press button then system will going to check the position automatically. (Please refer to System Status for the operate information) www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 20: Extraction

    2.5 Extraction (1) Plastic consumable set up: Insert the cartridges Push down each cartridges Insert Reaction Chambers Insert Tip Holder (ST cartridges with “click” (push down each wells) sound) www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 21 How to pull apart reagent cartridges strips: Slash open the dotted line with a cutter. The position of piercing pins (on the right) and filter tips (on the left). The 2 row in the middle should be EMPTY. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 22  You can load 1-12 strips on the tray depending on the number of samples you intend to process. (2) Start Program: (Please refer to System Status for operating information) Select the desired targets from homepage Select the Protocol then Press “NEXT” www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 23 Select Sample and Elution Volume then Press Select the Sample number then Press“NEXT” “NEXT” Edit the Sample ID then Press “NEXT” Edit the Elution ID then Press “NEXT” Scan Reagent Cartridge Barcode then Press “NEXT” www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 24 Zinexts do not guarantee with expired kit performance. “Admin” &“Service” ID have to take responsibility for kit usage status. System display a consumable Checklist, Tick corresponding boxes then Press “NEXT”. Note: The system will not allow user to go forward if not all consumable boxes are ticked. Add an Internal Control is optional.
  • Page 25: Cleaning & Maintaining The Magpurix 12 Evo

    3.0 Cleaning & Maintaining the MagPurix 12 EVO Once the sample or the buffer splash, stop to clean as following information. Caution: 3.1 Cleaning of Rubbish Tray and M motor To elevate M motor before cleaning of rubbish tray and M motor (1) Enter to setting and select “Function Test”...
  • Page 26 3.2 Cleaning of Piston (1) Enter to “setting” and select “System”, (2) ) Clean piston (Red Square) by kimwipes with select ”Piston” 75% ETOH and change new piston if there is significant impurity stain, salt crystal. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 27 3.3 Routine and Preventive Maintenance Two types of maintenance are to be performed on the MagPurix 12 EVO system instrument as listed in the table. For details on each type of maintenance, see below: Schedule Maintenance Type Performed by Routine Cleaning of sample Tray, M ...
  • Page 28 Clean the instrument body by removing dust gently with a dry, soft cloth. If the  outside of the MagPurix 12 EVO system is heavily soiled, or if any sample that may cause infection are adhering to the outside of the MagPurix 12 EVO system, wipe with a soft paper tissue, soaked with 75% ethanol.
  • Page 29: Troubleshooting

    Air leakage on the Filter Swap the air-leaked Tip with a new one liquid handling between Filter tips Blockage of tips and Too much starting 1. Decrease the amount of starting material. Use the www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 30 (≤ 20μl blood spot for forensics sample). DNA is sheared Bubbles form during To prevent bubble formation during mixing, make sure the or degraded mixing steps sample volume is at least the recommended volume listed www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 31: Specifications

    Input Power: Operating Temperature: 15-30 °C Operating Humidity: 30-80 % Fuse: F 5A 250 V UPS Specification On-Line, 1.5kVA/1200W, Battery energy (Watt-hours): 440 Wh EN 61326-1:2013 (IEC 61326-1:2012), EN 61326-2-6:2013 Safety IEC/EN61010-1, IEC/EN 61010-2-101 www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 32: Warranty

    Any consumption spare parts including fuses, light bulbs, UV lights, LED lighting sets. 2. For further information about the Warranty Policy and Extended Warranty offers, please contact your local representative / agent, or visit our website www.zinexts.com. www.zinexts.com MagPurix 12 EVO Instruction Manual...
  • Page 33: Revision History

    Zinexts Life Science Corporation www.zinexts.com 16F., No. 93, Sec. 1, Xintai 5th Rd., Xizhi Dist., New Taipei City 221416 , Taiwan (R.O.C) Tel: +886 2 2246 3579 Fax: +886 2 2243 8570 Mail: info@zinexts.com www.zinexts.com MagPurix 12 EVO Instruction Manual...