For Homecare Use:
Disinfect the Entire Circulaire® II Hybrid System After Each Treatment Day:
The Circulaire® II Hybrid Aerosol Drug Delivery System must be disassembled and
disinfected periodically in order to operate correctly. Westmed recommends full
disassembly and disinfection of all component parts of the entire system at the end
of each treatment day.
1.
Wash and dry your hands.
2.
Disassemble the entire Circulaire® II Hybrid System.
3.
Wash all parts (except the filter) in liquid dishwashing soap and rinse with sterile water
or soak all parts in a solution of 1 part household bleach and 50 parts water for 3 minutes.
4.
Place all parts on a clean surface (fresh paper towel) to air-dry. Place the Elastomeric
Reservoir on its flange to facilitate drainage.
5.
After the parts are dry, reassemble the Circulaire® II Hybrid System.
Please discuss the various methods of aerosol administration and proper cleaning
techniques with your medical provider or respiratory home care company.
CAUTION:
Do Not Use Vinegar. Do not reprocess, boil or microwave as high temperatures may
melt some of the plastic parts. Only the VixOne™ Nebulizer is top shelf dishwasher
safe.
When Your Care Is Critical
Westmed, Inc.
US Patent Numbers 5,020,530 & 5,613,489 and
5580 S. Nogales Highway
other US and International Patents Pending
Tucson, AZ 85706
©2013 Westmed, Inc. All rights reserved.
Toll Free: 800-975-7987
Phone: 520-294-7987
Assembled in Mexico
Fax: 520-294-6061
PN 78736 Rev. 2 2022/10
www.westmedinc.com
irculaire
®
C
II H
Aerosol Drug Delivery System
Instructions for Use
When Your Care Is Critical
ybrid
Need help?
Do you have a question about the Circulaire II Hybrid and is the answer not in the manual?
Questions and answers