Disinfect the Entire Circulaire
II Hybrid System After Each Treatment Day:
®
The Circulaire
®
II Hybrid Aerosol Drug Delivery System must be disassembled and
disinfected periodically in order to operate correctly. Westmed recommends full
disassembly and disinfection of all component parts of the entire system at the end
of each treatment day.
1.
Wash and dry your hands.
2.
Disassemble the entire Circulaire
3.
Wash all parts (except the filter) in liquid dishwashing soap and rinse with sterile water.
4.
Soak all parts in a solution of 1 part household bleach and 50 parts water
for 3 minutes.
5.
Place all parts on a clean surface (fresh paper towel) to air-dry. Place the Elastomeric
Reservoir on its flange to facilitate drainage.
6.
After the parts are dry, reassemble the Circulaire
Please discuss the various methods of aerosol administration and proper cleaning
techniques with your medical provider or respiratory home care company.
CAUTION:
Do Not Use Vinegar. Do not boil or microwave, as high temperatures may
melt some of the plastic parts. Only the VixOne™ Nebulizer is top shelf
dishwasher safe.
When Your Care Is Critical
Westmed, Inc.
MT Promedt Consulting GmbH
Altenhofstr. 80
5580 S. Nogales Highway
66386 St. Ingbert, Germany
Tucson, AZ 85706
Toll Free: 800-975-7987
Phone: 520-294-7987
US Patent Numbers 5,020,530 & 5,613,489 and
other US and International Patents Pending
Fax: 520-294-6061
©2013 Westmed, Inc. All rights reserved.
www.westmedinc.com
®
II Hybrid System.
®
II Hybrid System.
0 4 8 2
PN 75216 Rev .02
irculaire
II H
C
®
Aerosol Drug Delivery System
Instructions for Use
When Your Care Is Critical
ybrid

Advertisement

loading
Need help?

Need help?

Do you have a question about the Circulaire II Hybrid and is the answer not in the manual?

Questions and answers

Summary of Contents for Westmed Circulaire II Hybrid

  • Page 1 The Circulaire ® II Hybrid Aerosol Drug Delivery System must be disassembled and disinfected periodically in order to operate correctly. Westmed recommends full disassembly and disinfection of all component parts of the entire system at the end Instructions for Use of each treatment day.
  • Page 2 ybrid irculaire II H ® Aerosol Drug Delivery System Filling the Nebulizer: Variable Soft Expiratory Manifold Resistor Mouthpiece Wash and dry your hands. Detach the Nebulizer from the Manifold. Twist the purple cap counterclockwise until it releases and detaches from the Medication Bowl.