Customer Setup - Initial Velocity - Navitar 12X Zoom Instruction Manual

Motorized zoom systems for windows windows demonstration application
Hide thumbs Also See for 12X Zoom:
Table of Contents

Advertisement

Navitar Motorized Zoom Systems Instruction Manual for Windows
This configuration is stored within the controller's non-volatile memory and is
reloaded into the controller automatically when the controller power is turned on.
This configuration should only be edited with the controller connected and the
communications established, otherwise any changes will be lost. Distinct values
exist for each axis. The details below pertain to each axis equally.

8.6.1.1 Customer Setup - Initial Velocity

This value specifies the initial velocity, in steps per seconds, of the motor at start
up. From this value the motor will ramp to maximum velocity depending on the
Accel value. Generally, the heavier the load on the motor, the lower the initial
velocity value should be. If the initial velocity is set to high, the motor will not
start reliably and may stall during start up. In this case, lower the initial velocity.
If the initial velocity is set too low, the motor will take longer to reach its
destination.
Revision B
Recommended Initial Values
DC Encoded
Page 12 of 23
AXIS 1
AXIS 2
200
100
400
400
N/A
N/A
2/17/2004

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zoom 6000

Table of Contents