U AV - M I C R O P O R TA B L E A N A LY Z E R S | G L A 1 3 3 S E R I E S | U M/G L A 1 3 3 - E N R E V. B
. . .7
User Interface Operation
NOTE
You will no longer have internet access .
5 Establish communication with the analyzer using VNC:
a . Find the VNC icon shown in Figure 13 . (Check the Windows
Start men u application list) .
Figure 13 VNC Icon
b . Click on the VNC icon to execute the application; the
screen shown in Figure 14 appears .
Figure 14 VNC Get Started Screen
c . Enter the IP address of the analyzer . (Figure 7) The format
of the IP address is 192.168.1.XXX, where XXX consists of
three digits that vary from analyzer to analyzer .
d . When prompted, enter the password: lgrvnc .
The analyzer's software display should appear within the VNC
Viewer window .

Log In

To access the analyzer user interface features, log into the
system as follows:
1 Click the Security button on the Control Bar . (Figure 15)
Figure 15 Control Bar Security Button
2 For initial login, use the default Linux credentials for the
username and password (Figure 16), as follows:
User: lgr
Password: 3456789
WARNING!
If you change and forget this password, you will not be able
to recover it without a factory restore .
NOTE
There is only one Linux account .
Figure 16 Login Dialog Box
3 Click Login .