8. Firmware update procedure

If a firmware update becomes necessary, please follow the steps listed below.
Boot Version: 1.00.00.04
Main Version: 1.00.01.04
Firmware Erase:
1. Power off the device.
2. Press and hold down both arrow buttons and On/Off button.
3. The device will power up and start a long beep after 10 seconds.
4. Release all buttons.
Firmware Download:
1. Copy the update utility (WeberUsbFw.exe) and firmware file (WW100007.bin) to the same folder.
2. Press On/Off button.
3. The device will start in BOOT mode.
4. Connect USB cable to PC.
5. The PC will show a new device detected message, only at first time, and install drivers
automatically.
6. When done, run the update utility (WeberUsbFw.exe).
7. Click "Start". The utility and the device screen both will show the download progress.
8. When done, the device will show a success message.
9. Disconnect USB cable.
10. The device will power off.
11. Press On/Off button to turn on the device in normal mode.
Note: the device will again power on in BOOT mode if the firmware file is not compatible with the
device's serial number, or download process interrupted.
Retrieving Information:
In BOOT mode:
Press and hold down On/Off button at any time to show BOOT serial number and firmware revision.
In normal mode:
Power up the device while pressing and holding down the power button for 10 seconds. Serial number
and firmware revision will appear at the bottom of the screen.