Controller Diagnostics
In the unlikely event that the ROVI
will display one of several error codes to help diagnose the problem. Please see Table 1 for more
information.
Note: If performing the applicable procedure from Table 1 does not clear the corresponding fault,
DO NOT continue use of the power wheelchair. Turn off the power to the joystick controller and
consult your service provider immediately.
Table 1
BATTERY
GAUGE
An excessive voltage has been applied to the control system. This is usually caused by
10 bars flash
a poor battery connection.
The parking brakes have a bad connection. Check the parking brake and motor
9 bars flash
connections.
8 bars flash
A control system fault is indicated. Make sure that all connections are secure.
A joystick fault is indicated. Make sure that the joystick is in the center position be-
7 bars flash
fore switching on the control system.
6 bars flash
Battery charger connected.
5 bars flash
Right-hand motor wiring fault.
4 bars flash
Right-hand motor disconnected.
3 bars flash
Left-hand motor wiring fault.
2 bars flash
Left-hand motor disconnected.
1 bar flashes
Low battery voltage.
®