Advertisement

Symptom
Error code M005005
Error code M035000 or
M035005
Error code M046035
Error code M046120
Error code M046125
TROUBLESHOOTING AND REPAIR
TABLE 6-1
Possible Cause
START key held down too
long
Stuck START key
J4 not plugged in properly
Malfunctioning or damaged
J4 connector
Malfunctioning strain gauge
PCB
Excess drag in mechanical
drive assembly
Operator moved plunger
driver while pump was
infusing
Plunger driver moved too
quickly
Bent softpot bracket; Bad
softpot; Wiper spring out of
tolerance; Broken halfnut
components
Malfunctioning drive
assembly
LXXXXXX codes are issued by the Slave CPU.
MXXXXXX codes are issued by the Master CPU.
NOTE
6 - 5
Chapter 6
Solution
Operator error.
Replace keypad assembly.
Ensure proper connection of
J4.
Replace digital PCB.
Replace case/mechanical
drive assembly.
Replace case/mechanical
drive assembly.
Operator error.
Operator error.
Replace case/mechanical
drive assembly.
Replace case/mechanical
drive assembly.