Hadeco Smartdop 30EX Operation Manual
Hide thumbs Also See for Smartdop 30EX:
Table of Contents

Advertisement

GTIN: 00894912002098

Advertisement

Table of Contents
loading

Summary of Contents for Hadeco Smartdop 30EX

  • Page 1 GTIN: 00894912002098...
  • Page 2: Warnings & Cautions

    WARNINGS & CAUTIONS ® CAUTION Use the Smartdop 30EX under normal operating conditions, according to the instructions provided in this operation manual and under the operating conditions stated in the product specifications. CAUTION Federal law restricts this device to sale by or on the order of a licensed healthcare practitioner.
  • Page 3 WARNINGS & CAUTIONS CAUTION Do not use excessive force when disconnecting cables. 30EX accessories. ® CAUTION Use only designated Smartdop CAUTION The Doppler box should not have direct contact with the patient or be placed in the sterile area. The Doppler box cannot be sterilized or reprocessed. WARNING The Doppler box and noninvasive probes should be cleaned following any patient contact.
  • Page 4: Table Of Contents

    TABLE OF CONTENTS Warnings & Cautions SECTION 1 (Basic Operation) Introduction ........................1 Features ..........................1 1. Indications for Use, Contraindications, & Applications 1-1 Detection of Arterial and Venous Blood Flow Velocity ..........2 1-2 PPG & PV Arterial & Venous Studies..............2 2. Waveform Pattern Analysis 2-1 Doppler Arterial Waveforms .................3 2-2 Pulse Volume Arterial Waveforms .................4 3.
  • Page 5 TABLE OF CONTENTS 5-3-13 WAVEFORM ....................25 5-3-14 OTHERS-LANGUAGE...................26 5-3-15 OTHERS - PRB-KEY ..................26 5-3-16 OTHERS - CALIBRATION ................26 5-3-17 OTHERS - CAL PRINT (Printing Calibration)............27 5-3-18 OTHERS - BACKLIGHT ................27 5-3-19 OTHERS - AUTO OFF ...................27 5-3-20 OTHERS - RESET ..................28 5-3-21 OTHERS - DISCHARGE ................28 5-3-22 GUIDANCE ....................29 5-3-23 CURSOR ....................29...
  • Page 6 TABLE OF CONTENTS 13. Specifications ......................43 14.Operation, Storage & Transport Environments ...........44 15.System Includes .....................44 16.Cleaning ........................45 17.Disposal of Device ....................46 18.Warranty ........................46 19.Safety Standards ....................47-49 SECTION 2 (Testing Instructions) 1. Ankle Brachial Index ..................50-53 2. Segmental Blood Pressures ................54-56 3. Venous Compression Studies ................57 4.
  • Page 7: Section 1 (Basic Operation)

    Section 1 Section 1 Basic Operation...
  • Page 8: Introduction

    Smart-V-Link ® operation manual for instructions on downloading waveforms and performing tests with the Doppler and software. Manufactured by: Hadeco, Inc. 2-7-11 Arima, Miyamae-ku, Kawasaki, 216-0003 Japan FEATURES • Bidirectional Doppler with LCD, cuff inflator, & integrated printer Displays and prints velocity waveforms and numerical data...
  • Page 9: Indications For Use, Contraindications, & Applications

    Smartdop 30EX ® Basic Operation INDICATIONS FOR USE, CONTRAINDICATIONS, & APPLICATIONS INDICATIONS FOR USE The Smartdop® 30EX intended for use in the noninvasive evaluation of peripheral vascular pathology in patients by detecting arterial and venous blood flow in extremities. CONTRAINDICATIONS Not intended for use in ophthalmic or periorbital studies as the acoustic intensity is higher than the level recommended for use on and around the eye.
  • Page 10: Waveform Pattern Analysis

    Smartdop 30EX ® Basic Operation WAVEFORM PATTERN ANALYSIS DOPPLER ARTERIAL WAVEFORMS The Doppler arterial waveforms obtained from the lower extremity may be classified into six categories as an aid in interpretation. TYPE TYPICAL WAVEFORM FINDINGS The contour exhibits a steeply rising upslope at the onset of systole, rapid systolic downslope, and reverse flow (below baseline).
  • Page 11: Pulse Volume Arterial Waveforms

    Smartdop 30EX ® Basic Operation WAVEFORM PATTERN ANALYSIS PULSE VOLUME ARTERIAL WAVEFORMS A normal Pulse Volume arterial waveform will display a rapid rise in the upstroke during systole and a sharp peak at maximum amplitude, followed by a gradual downstroke following peak amplitude, and usually the presence of a dicrotic notch.
  • Page 12: Controls

    Smartdop 30EX ® Basic Operation CONTROLS FRONT VIEW 1) PRINTER To print waveforms, systolic pressures, ABI/TBI and velocity/pressure documentation. Outputs Doppler sounds. 2) SPEAKER To adjust sound volume. 3) VOLUME CONTROL For probe placement when not in use. 4) PROBE HOLDER 1 For probe placement when not in use.
  • Page 13: Operation Panel

    Smartdop 30EX ® Basic Operation CONTROLS OPERATION PANEL Indicates the mode selected. 1) MODE LEDs (Mode 1, Mode 2, Arm/Leg, Toe) Illuminates sequentially while the battery is charging. 2) MODE BUTTON To select the mode - Mode 1 (Arterial), Mode 2 (Venous), Arm/Leg Pressure, Toe Pressure.
  • Page 14 Smartdop 30EX ® Basic Operation CONTROLS To inflate the cuff or stop inflation and dump the blood 6) INF/DUMP (FREEZE) pressure cuff (Arm/Leg and Toe modes). BUTTON To freeze / unfreeze the waveform (measurement mode). Menu mode: Moves the cursor right and changes the mode or setting.
  • Page 15: Right Rear View & Probe

    Smartdop 30EX ® Basic Operation CONTROLS RIGHT REAR VIEW & PROBE To hold the paper roll in the printer. 1) ROLL SHAFT To connect the Doppler to a computer. 2) COMMUNICATION PORT To connect probe 1. 3) PROBE CONNECTOR 1 To connect probe 2.
  • Page 16: Setup & Basic Operation

    Smartdop 30EX ® Basic Operation SETUP & BASIC OPERATION TURNING THE UNIT ON / OFF Connect the probe with the round probe connection mark on the probe connector pointed up (12 o’clock). Press the power button to turn the unit Power ON: Probe connection mark Power OFF:...
  • Page 17: Checking The Battery Level

    Smartdop 30EX ® Basic Operation SETUP & BASIC OPERATION CHECKING THE BATTERY LEVEL Battery level indicator Turn the unit on and press the MENU button to display the Menu screen. The battery level indicator is shown in the upper right corner of the Menu screen.
  • Page 18: Installing The Printer Paper

    Smartdop 30EX ® Basic Operation SETUP & BASIC OPERATION INSTALLING THE PRINTER PAPER Turn the Smartdop ® 30EX ON and set the MODE to 1 or 2. Remove the paper roll shaft and set it into the paper roll. Insert and hold the edge of the paper into the paper entry slot located under the printer.
  • Page 19: Measuring Blood Velocity

    Smartdop 30EX ® Basic Operation SETUP & BASIC OPERATION MEASURING BLOOD VELOCITY This section explains the fundamentals of using the 30EX to measure blood velocity. For detailed ® Smartdop instructions on performing specific tests, see Section 2 Testing Instructions. ® Connect the probe to the Smartdop 30EX with the round probe connection mark on the probe...
  • Page 20 Smartdop 30EX ® Basic Operation SETUP & BASIC OPERATION MEASURING BLOOD VELOCITY (cont.) When the waveform becomes rhythmic and stable, wait 5 seconds or more without moving the probe on Mode 1 or 25 seconds or more on Mode 2. Then, press the INF/DUMP button or probe button (probe button function must be set to FREEZE or FRZ&PRINT) to freeze the waveform.
  • Page 21: Mode & Menu Settings

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS BASIC MODE SETTINGS The Smartdop® 30EX has the following 6 basic mode settings: Mode Type of Test How to Select Mode 1 Arterial (AC) Press the MODE button Blood Velocity (measurement mode only) Mode 2 Venous (DC) Press the MODE button...
  • Page 22: Menu

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU 5-3-1 MENU OPERATION Press the MENU button to enter Menu mode. The available menu options will appear on the LCD screen. Scroll through the available menu options using the buttons on the front control panel to navigate the menu.
  • Page 23: Menu For Blood Velocity Measurement Mode

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) Press the MENU button to go to Menu mode. The LED on the MENU button will be lit. 5-3-2 MENU FOR BLOOD VELOCITY MEASUREMENT MODE For measuring blood flow velocity waveforms and numerical data. Connect a 5, 8 or 10 MHz probe to select the menu.
  • Page 24: Menu For Blood Velocity Freeze Mode

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-3 MENU FOR BLOOD VELOCITY FREEZE MODE For observing blood flow velocity waveforms and numerical data. Press the probe button or the INF/DUMP button to select the menu; press again to exit. Menu Submenu Selections...
  • Page 25: Menu For Pressure Mode (Measurement & Freeze)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-4 MENU FOR PRESSURE MODE (MEASUREMENT & FREEZE) For measuring blood systolic pressures, performing Ankle Brachial Index (ABI) and Toe Brachial Index (TBI) studies and calculating ABI and TBI ratios. Press the ARM/LEG or TOE mode to select the menu.
  • Page 26: Menu For Ppg (Photoplethysmography)(Measurement & Freeze)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-5 MENU FOR PPG (PHOTOPLETHYSMOGRAPHY) (MEASUREMENT & FREEZE) For measuring and observing PPG waveforms. Connect the PPG probe to select measurement. Press the probe button or INF/DUMP button to select freeze; press again to exit. Menu Sub Menu Selections...
  • Page 27: Menu For Pv-Ac (Pulse Volume Arterial) Mode (Measurement & Freeze)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-6 MENU FOR PV-AC (PULSE VOLUME ARTERIAL) MODE (MEASUREMENT & FREEZE) For measuring and observing PV arterial waveforms and calculating PVR study results. Press the probe select button until both LED’s are lit. Press the INF/DUMP button to select freeze;...
  • Page 28: Memory-Store

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-8 MEMORY - STORE MEMORY-STORE is used to save waveform data into memory to view or print at a later time. MEMORY - STORE is available in Freeze mode only. See 5-2-1 “Measurement / Freeze Mode”...
  • Page 29: Memory-Read

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-9 MEMORY - READ MEMORY-READ is used to view previously stored waveform data. Press the MENU button to go to Menu mode. Scroll up (PRINT button) or down (LEG/TOE button) until MEMORY is highlighted.
  • Page 30: Dir (Flow Direction)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-10 MEMORY - CLEAR (cont.) indicate numbers which contain saved data. Scroll up (PRINT button) or down (LEG/TOE button) until the memory number you would like to clear is highlighted. Highlight “ALL”...
  • Page 31: Disp / Others-Disp (Waveforms/Data)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-12 DISP / OTHERS - DISP (Waveform/Data) DISP is used to change the LCD display from Waveform to Data. DISP is available in blood flow velocity measurement and freeze modes. When in measurement mode, DISP is a submenu of OTHERS in the menu.
  • Page 32: Waveform

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-13 WAVEFORM WAVEFORM is used to scroll between page 1 and page 2 of the waveform on the LCD in Blood Velocity and PV Arterial modes or to switch between the full and magnified view of the waveform in Pressure Mode.
  • Page 33: Others-Language

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-14 OTHERS - LANGUAGE LANGUAGE is used to change the LCD language preference. Press the MENU button to go to Menu mode. Scroll up (PRINT button) or down (LEG/TOE button) until OTHERS is highlighted.
  • Page 34: Others - Cal Print (Printing Calibration)

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-17 OTHERS - CAL PRINT (Printing Calibration) CAL PRINT is used to print the 4-step calibration after the velocity waveform. CAL PRINT is available in Mode 1 and Mode 2. Press the MENU button to go to Menu mode.
  • Page 35: Others - Reset

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-19 OTHERS - RESET RESET is used to reset the Doppler menu to the original default settings. Press the MENU button to go to Menu mode. Scroll up (PRINT button) or down (LEG/TOE button) until OTHERS is highlighted.
  • Page 36: Guidance

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-22 GUIDANCE GUIDANCE enables / disables testing instructions to appear on the LCD. It is available in ARM/ LEG and TOE pressure modes during measurement. Press the MENU button to go to Menu mode. Scroll up (PRINT button) or down (LEG/TOE button) until GUIDANCE is highlighted.
  • Page 37: P.clear

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS MENU (Cont.) 5-3-25 P.CLEAR P.CLEAR is used to clear the pressure data. It is available in ARM/LEG and TOE pressure modes. Press the MENU button to go to Menu mode. Scroll up (PRINT button) or down (LEG/TOE button) until P.CLEAR is highlighted.
  • Page 38: Shortcut Keys

    Smartdop 30EX ® Basic Operation MODE & MENU SETTINGS SHORTCUT KEYS Shortcut keys Functions Modes Reference section MODE + MENU Backlight ON/OFF All modes 5-3-18 MODE + RIGHT Move cursor right Pressure 5-3-23 MODE + LEFT Move cursor left Pressure 5-3-23 MODE + DOWN Zoom IN/OUT...
  • Page 39: Lcd Display

    Smartdop 30EX ® Basic Operation LCD DISPLAY BLOOD VELOCITY MODE 6-1-1 WAVEFORMS Smartdop 30EX automatically adjusts the baseline to best position. There are 4 ® baseline positions: bottom, 1/4 from the bottom, center, and 3/4 from the bottom. The waveform amplitude is automatically adjusted. The amplitude scale (velocity or frequency per division) is displayed in the upper left corner of the LCD.
  • Page 40: Pressure Modes (Arm/Leg And Toe Modes)

    Smartdop 30EX ® Basic Operation LCD DISPLAY PRESSURE MODES (ARM/LEG and TOE MODES) While in pressure mode, Smartdop® 30EX automatically calculates arm, leg and toe systolic pressures. In addition, Smartdop® 30EX automatically calculates and prints ABI and TBI ratios. See Section 2 of this manual for detailed instructions on using the pressure mode to perform ABI, TBI and Segmental Pressure studies.
  • Page 41: Error Messages

    Smartdop 30EX ® Basic Operation LCD DISPLAY PRESSURE MODES (ARM/LEG and TOE MODES) (Cont.) 6-2-2 ERROR MESSAGES If an error occurs during blood pressure measurement, one of the following error messages will appear on the LCD. (Reserved) Waveform error Pressure exceeded 300 mmHg Systolic pressure not found Exceeded time limit Cuff pressure did not increase...
  • Page 42: Printing Waveforms & Data

    Smartdop 30EX ® Basic Operation PRINTING WAVEFORMS & DATA HOW TO PRINT Waveforms can be printed while in Freeze mode only. Press the PRINT button. NOTE Pressing the probe button also activates the printer when the probe button function is set to either PRINT or FRZ&PRINT.
  • Page 43: Print Samples

    Smartdop 30EX ® Basic Operation PRINTING WAVEFORMS & DATA PRINT SAMPLES Mode 1 (Arterial) Time scale 0.2 sec/div. (approx. 25 mm/s) Combined forward and reverse 0.2 sec. 5 sec. max. Mode 2 (Venous) Time scale 1 sec/div. (approx. 5 mm/s) Separation of arterial and venous flow velocity 1 sec.
  • Page 44 Smartdop 30EX ® Basic Operation PRINTING WAVEFORMS & DATA PRINT SAMPLES (Cont.) Arm/Leg and Toe Pressure Modes (press print button less than 1 second). Arm only Leg only Toe only Arm & Leg Arm & Toe Arm, Leg, & Toe (Save the arm pressure last) Arm/Leg and Toe Pressure Modes (press print button longer than 2 seconds).
  • Page 45: Numerical Velocity

    Smartdop 30EX ® Basic Operation PRINTING WAVEFORMS & DATA NUMERICAL VELOCITY Mode 1 (Arterial) Parameters Abbrs. Systolic velocity [cm/s] Mean velocity [cm/s] Heart rate [bpm] Mode 2 (Venous) Parameters Abbrs. Systolic velocity [cm/s] Mean velocity [cm/s] NOTE Computations are based on the following parameters. If the heart rate cannot be calculated because of an over +/-25% irregularity due to Arrhythmia or a venous waveform, “HR***”...
  • Page 46: External Outputs

    Smartdop 30EX ® Basic Operation EXTERNAL OUTPUTS HEADSET When using an external headset (not provided), sound to the speaker will be cut off. COMMUNICATION PORT (3.5 mm Jack) To observe the waveform in high resolution on a PC monitor or to store the waveform and numerical data into a computer for future reference, or to create a report of standardized testing modules.
  • Page 47: Symbol List

    Smartdop 30EX ® Basic Operation 10. SYMBOL LIST Symbols Descriptions Symbols Descriptions Type BF applied part Freeze Headset Mode button USB connector Print button AC adaptor connector Menu button Probe connector Arm button Caution Leg/Toe button Power ON Probe selection button Federal law restricts Power Off...
  • Page 48: Operating Principles

    Smartdop 30EX ® Basic Operation 11. OPERATING PRINCIPLES 30EX bidirectional Doppler is designed to obtain blood flow velocities through ® The Smartdop the use of ultrasound, which is transmitted from the probe to the patient’s body and reflected by the blood (hemocyte, etc.). The unit amplifies the high frequency oscillation output and then supplies it to the transmitter transducer.
  • Page 49: Block Diagram

    Smartdop 30EX ® Basic Operation 12. BLOCK DIAGRAM...
  • Page 50: Specifications

    Smartdop 30EX ® Basic Operation 13. SPECIFICATIONS Freq. Model# GTIN# Ispta* (in situ) [mW/cm²] Probes (For transcutaneous use): 5 MHz BT5M05S8A 00894912002333 720 mW/cm² or less 8 MHz BT8M05S8A 00894912002357 720 mW/cm² or less 10 MHz BT10M5S8A 00894912002302 720 mW/cm² or less *Ispta: Spatial Peak - Temporal Average Intensity Power: NiMH rechargeable battery pack or AC adaptor...
  • Page 51: Operation, Storage & Transport Environments

    Smartdop 30EX ® Basic Operation 14. OPERATION, STORAGE & TRANSPORT OPERATING ENVIRONMENT • 10 to 37°C • 85% humidity or less with no condensation CAUTION Do not drop or place in an area where the unit may fall. CAUTION Do not use simultaneously with an electrocautery defibrillator, other ultrasonic device or a mobile phone.
  • Page 52: Cleaning

    Smartdop 30EX ® Basic Operation 16. CLEANING Clean the Doppler & probes after use. DOPPLER BOX To clean the Doppler box, wipe down with a damp cloth, then with a soft dry cloth. Take care not to allow water to penetrate the unit. CAUTION The Doppler box should not have direct contact with the patient or be placed in the sterile area.
  • Page 53: Disposal Of Device

    Smartdop 30EX ® Basic Operation 17. DISPOSAL OF DEVICE Please follow local ordinances or regulations for proper disposal of the device, accessories & packaging. Do not place the battery in fire. Doing so may cause an explosion and/or injury. 18. WARRANTY The Doppler box is guaranteed for a period of 2 years from the date of purchase when used under normal conditions.
  • Page 54: Safety Standards

    Smartdop 30EX ® Basic Operation 19. SAFETY STANDARDS The Smartdop ® 30EX conforms to the following standards: Manufacturing standard: IEC60601-1 Protection class against electric shock : Class II device : Internally powered equipment Protection grade against electric shock : Type BF applied part Leak current: Based on IEC60601-1 Items...
  • Page 55 Smartdop 30EX ® Basic Operation Guidance and Manufacturer’s Declarations (cont.) Electromagnetic Immunity 30EX is intended for use in the electromagnetic environment specified below. The user of The Smartdop ® 30EX should assure that it is used in such an environment. ®...
  • Page 56 Smartdop 30EX ® Basic Operation Guidance and Manufacturer’s Declarations (cont.) Electromagnetic Immunity (cont.) 30EX is intended for use in the electromagnetic environment specified below. The user of The Smartdop ® 30EX should assure that it is used in such an environment. the Smartdop ®...
  • Page 57 Section 2 Section 2 Testing Instructions...
  • Page 58: Ankle Brachial Index

    Smartdop 30EX ® Vascular Testing Instructions ANKLE BRACHIAL INDEX (ABI) The Ankle Brachial Index is the ratio of the highest systolic pressure at the arm to the systolic pressure at the ankle. This simple test is performed much like taking a standard blood pressure, except that a Doppler is used instead of a stethoscope.
  • Page 59 Smartdop 30EX ® Vascular Testing Instructions ANKLE BRACHIAL INDEX (ABI) (Cont.) To see a magnified view of the waveform, press and hold the MODE button, then press the LEG/TOE button. Press again to return the full waveform view. Or, use the WAVEFORM option in the menu.
  • Page 60 Smartdop 30EX ® Vascular Testing Instructions ANKLE BRACHIAL INDEX (ABI) (Cont.) Apply ultrasound gel to the posterior tibial artery and place the Doppler probe on the skin. Hold the probe on the posterior tibial artery at a 45 to 60 degree angle against flow (see figure C). Adjust the probe angle until best sounds are heard and a steady waveform appears on the LCD.
  • Page 61 Smartdop 30EX ® Vascular Testing Instructions ANKLE BRACHIAL INDEX (ABI) (Cont.) Calculating the ABI To calculate the Ankle Brachial Index ratio, divide the ankle pressure by the highest arm pressure or refer to the charts on the Pressure Index Guide provided with your Doppler. Ankle systolic pressure ABI = Arm systolic pressure...
  • Page 62 Smartdop 30EX ® Vascular Testing Instructions SEGMENTAL BLOOD PRESSURES Segmental pressure studies are performed with the same method as an Ankle Brachial Index, but incorporate additional cuffs wrapped above and below the knee, and optionally at the high thigh, in additional to the ankle.
  • Page 63 Smartdop 30EX ® Vascular Testing Instructions SEGMENTAL BLOOD PRESSURES (Cont.) When the waveform is steady and the INF/DUMP LED begins to blink, hold the probe in place and press the INF/DUMP or probe button to activate the automatic cuff inflator. The Smartdop ®...
  • Page 64: Segmental Blood Pressures

    Smartdop 30EX ® Vascular Testing Instructions SEGMENTAL BLOOD PRESSURES (Cont.) Interpreting the Results Observe pressure differences between adjacent cuff sites on the same leg. Pressure differences between cuff sites are used to localize the disease. Pressure differences between two adjacent levels of less than 20 mmHg is considered normal within limits.
  • Page 65 Smartdop 30EX ® Vascular Testing Instructions VENOUS COMPRESSION STUDIES Venous compressions are auditory tests consisting of 2 maneuvers. These are, compression of the metatarsal arch to verify augmentation followed by compression of the calf muscle to verify valvular competence. Venous compressions are performed to detect enlarged veins due to the presence of malfunctioning valves.
  • Page 66: Ppg Toe Pressures & Toe Brachial Index (Tbi)

    Smartdop 30EX ® Basic Operation PPG TOE PRESSURES & TOE BRACHIAL INDEX (TBI) Requires the optional PPG (Photoplethysmography) probe {PG-21}, supplied separately. PPG (Photoplethysmography) toe pressure studies and related TBI’s (Toe Brachial Index) are performed to assess the vascular condition of the foot by obtaining systolic pressures at the toe.
  • Page 67 Smartdop 30EX ® Vascular Testing Instructions PPG TOE PRESSURES & TOE BRACHIAL INDEX (TBI) (Cont.) T ake the systolic pressure at the toe After taking the arm pressures and documenting the PPG waveform, press the MODE button to select Toe pressure mode (the TOE LED will be lit).
  • Page 68: Ppg Venous Reflux

    Smartdop 30EX ® Vascular Testing Instructions PPG TOE PRESSURES & TOE BRACHIAL INDEX (TBI) (Cont.) Press the INF/DUMP button to return to the monitoring screen and repeat the test on the other foot. NOTE If you would like to print the pressure results, before pressing INF/DUMP to return to the monitoring screen, press the PRINT button less than 1 second to print numerical data only or longer than 2 sec-...
  • Page 69 Smartdop 30EX ® Vascular Testing Instructions VENOUS REFLUX Requires the optional PPG (Photoplethysmography) probe {PG-21}, supplied separately. Venous reflux studies use a PPG probe affixed adjacent to the Posterior Tibial artery. The patient is asked to dorsiflex the foot, which acts as a pump for venous augmentation. Recovery time is the time required for refilling of the vein to occur.
  • Page 70: Pulse Volume Recordings (Pvrs)

    Smartdop 30EX ® Vascular Testing Instructions VENOUS REFLUX (Cont.) Allow the sensor to equilibrate. Then, press the PRINT button or the probe button to begin the test. Ask the patient to dorsiflex the foot the specified number of times, following the foot animation on the LCD.
  • Page 71 Smartdop 30EX ® Vascular Testing Instructions PULSE VOLUME RECORDINGS (PVRs) PVRs use pulse volume (pneumoplethysmography) to identify changes in leg blood volume. Rather than temporarily occluding blood flow, the cuffs are only inflated slightly so that blood continues to flow. Blood pulses in the arteries cause the vessels to expand which changes the air volume in the cuff.
  • Page 72 Distributed by: Koven Technology, Inc. 477 N. Lindbergh Blvd., Suite 220 St. Louis, MO 63141 Tel: 800-521-8342 • Fax: 314-542-6020 E-mail: info@koven.com www.koven.com GLN: 0894912002005 OM-1326 rev5 Smartdop 30EX Operation Manual Copyright © Koven Technology, Inc.