User Manual I. Disclaimer & Warranty • The intellectual property rights for this manual belong to Holland Green Science Corporation (hereinafter referred to as “the Company”). • Products of the company are under the patent protection of China and other countries and regions (inclusive of obtained patents and patents currently under application).
Page 4
User Manual will provide Buyer with service data and/or a Return Material Authorization ("RMA"), which may include biohazard decontamination procedures and other product-specific handling instructions, then, if applicable, Buyer may return the defective Products to the Company with all costs prepaid by Buyer.
User Manual OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. THE COMPANY DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. NOTHING IN THIS MANUAL OR IN THE TERMS AND AGREEMENT AGREED TO AT THE TIME OF PURCHASE SHALL BE CONSTRUED TO PROVIDE FOR THE TRANSFER, ASSIGNMENT OR ALIENATION OF THE LIMITED WARRANTIES PROVIDED BY THE...
Page 6
User Manual BREACH OF CONTRACT, TORT, INDEMNIFICATION, OR OTHERWISE, BUT EXCLUDING LIABILITY OF THE COMPANY FOR BREACH OF WARRANTY (THE SOLE REMEDY FOR WHICH SHALL BE AS PROVIDED UNDER THIS SECTION) SHALL NOT EXCEED AN AMOUNT EQUAL TO THE LESSER OF (A) THE TOTAL PURCHASE PRICE THERETOFORE PAID BY BUYER TO THE COMPANY WITH RESPECT TO THE PRODUCT(S) GIVING RISE TO SUCH LIABILITY OR (B) ONE HUNDRED THOUSAND DOLLARS ($100,000), WHICHEVER IS LESS.
User Manual Safety IMPORTANT Safety Items to be Observed. Safety Warnings ! Dangerous (may result in serious loss of property or casualties) The instrument must be grounded and away from sources of electromagnetic interference. Before use, please confirm the voltage and frequency of the power supply match the instruments’...
User Manual Ygrasia Moisture Analyzer – Introduction III. The Ygrasia moisture analyzer provides reliable, fast, and accurate measurements of trace levels of moisture content within your samples. The moisture analyzer uses the Loss On Drying (LOD) method to measure moisture. The moisture analyzer first weighs a sample and then heats it until dry and weighs it a second time when dry.
Page 9
User Manual After Sales Support If problems are encountered or technical assistance is required, when installing or using the instrument, please contact the service department at serviceusa@hollandgreenscience.com. The Company may provide technical assistance and information regarding the instrument or equipment or service without charge at its sole discretion. Buyer assumes sole responsibility for any reliance on or use of such assistance and information, and the Company makes no warranty thereon.
User Manual IV. Technical Specifications Product Name Ygrasia 110 Parameters Model 10813001 Capacity 110g Resolution 0.01 g 5 ℃ ~ 35 ℃ Operating Temperature Display LCD (White backlight with black front) Calibration External Calibration Interface RS232 Pan Size ɸ90mm Packing Size 490 x 350 x 360 mm Gross Weight 7.5kg...
User Manual Inspection Packing List Unpack the equipment carefully and check for any damage which may have arisen during transport. In the event of identified damage, please contact serviceusa@hollandgreenscience.com. • Carefully unpack the shipping container. • Check the instrument and accessories against the ‘Packing List’ below. The package includes the following items: Item Description.
User Manual VII Installation instructions 7.1 Precautions • The moisture analyzer is used to measure the moisture content of a sample. Improper operation of the analyzer may result in personal injury and/or equipment damage. • Make sure the power cord is connected according to the instructions in the manual. •...
User Manual Installation of accessories Install the accessories according to Figure x. Installation order from top to bottom: 1-2-3-4-5 1. Aluminum scale pan 2. Triangle scale frame 3. Scale pan brackets 4. Wind cover 5. Stainless steel plate Other installation requirements •...
User Manual VII Operating Instructions 8.1 Display 8.2 Icon Introduction • The percentage of dry weight • The percentage of moisture Gram • • Temperature • Set time and heat time • Heating mode • Standard mode • Gentle heating mode •...
User Manual 8.3 Keypad NAME Key features ON/OFF Power on/off, off the internal DC power supply START/STOP Start heating/Stop heating TARE Weight cleared, settings exit MENU Enter the setting menu CONVERSION Switching measurement results ENTER Confirm the current settings, go to next RESET Return to the weighing mode(short press), calibration balance(long press)
User Manual 8.4 Testing for the first time • When using the moisture analyzer for the first time, the settings are set to factory defaults. Change factory settings if necessary. • Put an empty sample pan on the holder (see section 7.2). •...
User Manual 8.5.2 Heating mode and setting • Standard heating mode :This mode is the factory default setting. It is suitable for most samples. The heating temperature can be set. The sample is heated to the set temperature and remains at the temperature until the test ends automatically. •...
User Manual setting (time setting). The icon will display the current temperature when its heating. 8.5.5 Heating time setting • The operator needs to set the stop time when in timing stop mode. Press the “MENU” to enter the parameter setting mode and press the “ENTER” key until the flashing.
User Manual IX Print Test Results • Please choose a printer that is compatible with the moisture analyzer • Use the following steps to print the moisture analyzer measurements: 1. Press and hold "Menu" until the display shows "print". 2. Select the print language using the “up” or “down” keys: L-C (Chinese) or L-E (English).