2.0 InTROdUCTIOn The OPTIKA zoom stereo microscopes SZP series are high performance Galileian Common Main Objective stereo microscopes with continuous zoom system. Total magnification varies from ,4x to 30x depending on the eyepieces and additional lenses used. It is the best equipment to examine all types of gross materials in three dimensions for industrial, biological and educational applications.
3.0 USeR nOTICe Safety Precaution Carefully open the box, avoid the accessories, like lens, dropping to ground and being dam- aged. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment. Make sure the stage is smooth, horizontal and firm enough. To prevent the microscope body from turning over, its pivot angle must be limited to 30°.
4.0 ASSeMBLY The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly. When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. Point 3: insert the head into the holder of your stand.
4.0 ASSeMBLY detailed Assembly Procedure 4.2.1 Mounting binocular viewing head (Fig.1, Fig.2) 1. Rotate the objective (1) into the objective mount thread on the lower end of the zoom body (2), as (3) in Fig.. 2. Aligning the positioning pin of the zoom body (3) to the positioning groove on the binocular viewing head (4), insert the dovetail mount at the bottom of the viewing head into the mounting port on the...
Page 8
4.0 ASSeMBLY 4.2.2 Mounting binocular head Insert the binocular head and zoom body assembly into the bracket of your stand. 4.2.3 Mounting eyepieces (Fig.4) 1. Remove the eyepiece dust caps and gently in- sert two adjustable eyepieces (10) into the eye- piece sleeves until they stops, as in Fig.4.
6.0 MAInTenAnCe 5.0 AdJUSTMenT And OPeRATIOn Adjusting Binocular Viewing Head 5.1.1 Adjusting interpupillary distance (Fig.10) Different users have different interpupillary distanc- es. So when users change, do the interpupillary distance adjustment. While looking through the eyepieces, hold the left and right of the binocular assembly (1) and adjust the eyepieces by opening or closing them for bin- ocular vision until the left and right fields of view coincide completely.
5.0 AdJUSTMenT And OPeRATIOn 6.0 PHOTOGRAPHY And VIdeO ATTACHMenT Photography and Video Attachment Assembly (Fig.16 ~ Fig.19) ● The optional light splitter unit (3) makes it possible to perform photography or video observation. 1. Loose the bolt (1) with provided hex wrench and re- move the binocular head (4) away from the zoom body. 2.
X Magnification of zoombody X Magnification of adapter X 1.5 Switching the Light Path Pull out the light switching lever (10) to open the light path, as in figure 19. Light flux ratio: Binocular: photography=8:. 7.0 SPeCIFICATIOn model no. SZP-6 SZP-8 SZP-10 ParameterS Zoom ratio 1:10 oBjective magniFication 0.8×...
8.0 TROUBLeSHOOTInG Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact your local franchiser for assistance.
9.0 ReCOVeRY And ReCYCLInG Art.13 Dlsg 5 july 005 N°151. “According to directives 00/95/EC, 00/96/EC and 003/108/EC relating to the reduction in the use of hazardous substances in electrical and electronic equipment and waste disposal.” The basket symbol on equipment or on its box indicates that the product at the end of its useful life should be collected separately from other waste.