oticon REM AutoFit Quick Manual page 6

With verifit link

Advertisement

6
Manual: lets you fine-tune the instrument gain to get even closer to target
or to personalize the fitting based on your client's comments. This manual
process can be done after or instead of the automatic process.
Within the REM AutoFit tool, use the fine-tuning controls to adjust the
gain and then manually run a measurement to verify the target match.
a.
Select the desired measurement level(s) and MPO, if necessary.
Measurements take place in a sequence for all selected levels starting
with the softest. We recommend that you run MPO as a separate mea-
surement so you can prepare the client for louder levels.
b.
If performing on-ear measurements, instruct your client to remain still
during each sequence of measurements.
Click Start on the relevant side to measure one side. Click Both to
c.
measure both sides (binaurally). Binaural measurements take place
simultaneously if using Genie 2 | 2019.2 and Verifit2 with software
version 4.18 or later. Otherwise they take place sequentially.
d.
When doing simultaneous binaural measurements, if the signal reaching
left and right reference microphones is not equal, e.g. due to the client
not facing the speaker, the measurement will be paused and Verifit2's
Binaural Sound Field Assist tool will show instructions on the Verifit2
screen to help balance the signal reaching left and right.
e.
The system automatically turns off the signal when the measurements
are complete. If you need to stop the measurement prior to completion,
click Stop.
f.
If you are satisfied with the measurements, proceed to step 3: Finish.
Use the buttons between the graphs to change the graph view to show/hide
targets, percentiles, LTASS, different levels and SII.