Settings in other tools
The following controls are set automatically when REM AutoFit with VerifitLINK
is open and restored to their original states afterwards:
Adaptation Manager
Noise reduction
Feedback Analyzer
Directionality setting
Speech Rescue
3
Step 3
Current setting in OpenSound Navigator™
On
Pinna Omni. For styles with a single
microphone, directionality is not applicable.
Off
Step-by-step Guide
On the Fitting step in Genie 2, enter the REM tool from the More Tools menu
in the left taskpane. Launch REM AutoFit with VerifitLINK by clicking the
VerifitLINK button at the bottom of the REM tool.
Step 1: Setup
1.
Select a Measurement method from the drop-down list:
On ear or Test box
2.
Select a Fitting rationale. You can use all available fitting rationales
on the Verifit, except Camfit.