3 Years Warranty www.bemergroup.com bemer-set pro :: User Manual en – bemer system Only valid for USA november 2020...
Page 2
Congratulations on the purchase of your new BEMER set. These high quality products are manufactured by us with the greatest of care and meet our high quality standards, as well as the European and international production standards of medical products.
Contents Scope of Delivery Applications – B.BOX Professional Foreword Default Settings – using the B.BOX Professional Safety Instructions Updates Intended Use B.SCAN Contraindications Wall Mount Assembly Precautions Rechargeable battery – B.BOX User Information Cleaning Basic Plan Maintenance General Information Proper Disposal Examples of Application Troubleshooting Connections –...
Foreword Symbols found in the operating instructions This instruction manual belongs to this device. It contains important informa- Warning / Caution – failure to observe tion for usage and handling. Read these instructions completely. Non-obser- can result in possible injuries or property vance of these instructions can lead to injury of the user or damage to the unit.
Application part type Application part is type BF • Consult with your doctor before using BEMER devices, if you are in the care Serial number of a doctor. • Before you connect the BEMER device to the power supply, please make sure Refer to instruction manual / booklet that the power supply is identical with that of the B.BOX power pack supply.
Page 8
BEMER devices are not toys. Keep pets away from devices. the airway or cause difficulty in breathing. • The BEMER device is not intended to be used for use by • Stimulation should not be applied transthoracically (through people under 18 years of age.
• Other conditions often requiring immunosuppressive therapy, e.g. autoimmune diseases or dermatological diseases are not contraindications to the use of BEMER therapy. BEMER therapy application has to be cleared by physician in charge. • Do not use the device if you have a diagnosed Deep Vein Thrombosis (DVT) •...
Page 10
With the following conditions, it is strongly recommended Use of the device beyond that which has been that you consult a medical doctor prior to the use of BEMER specified is considered contrary to the intended therapy: purpose.
Page 11
BEMER therapy has to be discussed with the treating medical doctor. • If you feel any discomfort while using BEMER immediatley stop all applications and consult your physician in charge. Further applications of BEMER have to be cleared by him or her.
Over areas of the skin which lack normal sensation. medical doctor is advised to contact BEMER medical support • In general, application of BEMER therapy in users with active electronic implants and devices (such as medications pumps, neurostimulators etc) has to be discussed and cleared by the...
Page 13
Warning Use of this equipment adjacent to or stacked with other equipment should be avoided be- cause it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to veri- fy that they are operating normally. Warning Use of accessories, transducers and cables other than those specified or provided by...
Velcro fastener strap to nearly every conceivable part of the body. When using on disabled or mobility impaired people, the B.PAD The BEMER devices may be used for the following: • To reduce stress can provide a particularly valuable service.
Page 15
P1-P3 additionaly to the basic The existing, predefined programs cover the wide range of Plan. Overdosing or habituation effects from using the BEMER application requirements needed for daily use. system are not known.
This The effects of the specific signal configuration are is the basis of treatment with BEMER therapy and defines the scientifically proven and essentially independent standard of treatment for conditions of reduced well-being.
Page 17
cycle 1 week 1 week 2 week 3 week 4 week 5 week 6 morning Level 1 plus Level 2 plus Level 3 plus Level 4 plus Level 5 plus Level 6 plus evening Level 1 Level 2 Level 3 Level 4 Level 5 Level 6...
Page 18
Strong intensity for muscle stimulation 20 min. and training BEMER Usage Support P3 for active athletes If you still have questions or are uncertain, please contact your treating medical doctor or contact For local application using the pre-defined programs, the BEMER medical support: “HIGH”...
General Information Both the Basic Plan and all other application sug- The Basic Plan treatment is carried out using the B.PAD ac- gestions are based on many years of experience. cording to the basic plan using intensities 1-6 (short applica- This experience has also proven that regular ap- tion).
Examples of Application B.SCAN The B.BOX has an integrated test device for checking the performance of your application module. B.PAD Handy, flexible and universally mobile. This application module is made for local treatment.
Page 21
B.GRIP Multi-functional grip for B.SPOT. B.SPOT Wall mount Flexible, portable application module for local treatment. A wall mount for easy fixation of the B.BOX – for example, on the wall next to a double bed.
Connections for the applicator modules (A1 and A2) Basic position, recommended Position 1 and position 2, recommended when using the wall mount for table application with different angles. Only use approved BEMER accessories. Suitable for software updates and approved BEMER accessories.
Start-up Procedure – B.BOX The plug on the power cord is the AC mains disconnect device and must remain readily operable. To completely disconnect this apparatus from the AC mains, disconnect the power supply cord plug from the AC receptacle. Turn the B.BOX over and connect it to the adapter.
Display – B.BOX Professional Wait several seconds after turning on the device unit until it is ready for operation 2 3 4 (see display below). A1 08:00 20 : 10 From the main window you can access the individual menu items by pressing the Intensity Mode various buttons.
Page 25
Applicators (display the remaining Displays the active mode (the intensity time of the ongoing application) and status of the plus signal) Application window (displays the application Power Save (default activated) progress incl. level and period of time) Memory card (if inserted) Applicator window (displays the connected application module on the chosen port) Volume (in current setting)
Applications – B.BOX Professional By pressing the buttons [A1 / A2] you can Example Short application (intensity levels 1-10) switch between the connected application • A B.PAD is connected on port [A1] Press the button [INTENSITY]. modules. • and a B.SPOT on port [A2]. Select the desired intensity by pressing •...
Page 27
Example – short application with intensity 1 (using the application B.PAD via port A1) B . PAD Program selection Example – Program P1 (with the application module B.PAD via port A1) • Press the button [PROGRAM]; if a memory card is inserted, press the button [PROGRAM] again.
Default Settings – application parameters / Basic settings – B.BOX Professional The parameters [PLUS] and [INTENSITY] Setting the signal volume can be adjusted to be the default setting. • Press the [SET] button. • Press the [VOLUME] button. Setting the plus Signal as default •...
Page 29
Adjusting the display brightness Setting the time To access the system information • • • Press the [SET] button. Press the [SET] button. Press the [SET] button. • • • Press the [BRIGHTNESS] button. Press the [TIME] button. Press the [INFO] button. •...
Page 30
The Power Save mode influences the stand-by characteristics Activate/deactivate Power Save mode of your machine. If it is deactivated, the machine turns itself off (default activated) after approx. one hour of inactivity (switch-off time 2 minutes). • Press the [SET] button. •...
Updates BEMER product are a secure investment and are therefore designed to transfer product advancements with the corresponding updates to the B.BOX through the communication or maintenance interface. Because different approaches are possible, you will receive exact instructions for proceeding with any updates along with the delivery.
B.SCAN The B.SCAN is integrated in your B.BOX and is designed to check your BEMER product along with safety precautions for their proper operation. The B.SCAN makes the specific BEMER Signal configuration audible. The device is switched on by pressing the button once; pressing and holding the button down (for several seconds) will switch off the device.
Wall Mount Assembly – Pro-Set Assemble the wall mount as shown. Please note the course of the cable duct and use to assemble according to the circumstances. Basic position recommended when using the wall mount. Position 1 and position 2, recom- mended for table application at different angles.
The high capacity, rechargeable battery used with the B.BOX makes it indepen- dent of electricity obtained from a power outlet and allows mobile application of the BEMER technology. • Completely charge the battery before using for the first time. •...
Page 35
Assembly / Changing the battery • Loosen the screws and remove the cover of the battery compartment. • Tilt the rechargeable battery into the battery compartment and make sure that it is securely in place. • Now attach the battery.
• B.BOX // B.GRIP // B.SPOT Repeat the cleaning procedure at regular intervals. • BEMER devices and the associated components are neither water- Observe the safety notes. proof nor protected against spray water. For this reason, the housing • Do not use abrasive, alkaline, or acidic cleaning agents.
Batteries do not belong in household garbage! • Before you send a BEMER product for repair to the manufacturer or to one of their authorized specialists, make sure to first carry out a generalized error trouble shooting check.
Troubleshooting General information BEMER devices are equipped with intelligent electronics. There- If an error still persists or if one has not been properly located, fore they can automatically detect and display errors. Should an please follow the instructions of the corresponding error code error be displayed or should you suspect a malfunction, then or contact your responsible service address.
Page 39
ERROR MESSAGE CAUSE TROUBLESHOOTING ERROR 1 An active applicator has been removed during the Reconnect the applicator and start the application again. application. ERROR 2 The temperature of the product is too high. Check the environmental conditions and let the product cool down.
Technical Data B.GRIP Art. No. 431000 B.BOX Pro Art. No. 420201 Dimensions (W x H x D) 12 x 12 x 5 cm Dimensions (W x H x D) 22 x 23 x 7 cm Weight 200 g Weight 1.4 kg Connection plug 6 pin Operating voltage...
Page 41
Attachment belt (B.GRIP) Art. No. 450700 B.BOX power supply Art. No. 440100 Dimensions (W x H x D) 730 x 120 x 4 mm Primary voltage 100-240 V AC Weight 50 g Primary current Surface material Neoprene Frequency 50-60 Hz Secondary voltage 15 V DC Vehicle connection cable...
Page 42
Technical Specifications BEMER PRIMARY MODE OF ACTION Non-invasive tissue stimulation via magnetic field induction Waveform Pulsed asymmetric, constant amplitude during treatment Shape Sinusoidal, monopolar Pulse repetition rate All accessories: 10-30 Hz Single pulse duration All accessories: 10 - 33 µs Maximum Power density applied: All accessories: 35 - 150 µT...
Page 43
Environmental conditions during operation Maximum temperatures Temperature +41° - +104°F B.SPOT 108°F Air pressure 700 … 1060 hPa 3000 m (ASL) Relative humidity 15 … 93 % RH, +158 °F non-condensing Standard marking IEC 60601-1 Medical electronic equipment Storage and transportation IEC 60601-1-2 Electromagnetic compatibility Temperature...
Guideline and Manufacturer‘s Declaration of Conformity (tables according to IEC 60601-1-2) Electromagnetic emission The BEMER-SET PRO is designed for operation in the electromagnetic environment listed below. The customer or user of the BEMER-SET PRO should ensure that the product be used in such an environment. EMISSION TESTS COMPLIANCE ELECTROMAGNETIC ENVIRONMENT –...
Page 45
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the BEMER Professional device, including cables specified by the manu- facturer. Otherwise, degradation of the performance of this equipment could result.
Page 46
Electromagnetic immunity The BEMER-SET PRO is designed for operation in the electromagnetic environment listed below. The customer or user of the BEMER-SET PRO should ensure that the product is used in such an environment. IMMUNITY TESTS IEC 60601 TEST LEVEL...
Page 47
Electromagnetic immunity The BEMER-SET PRO is intended for use in the home health care (e.g. domiciles, homes) or professional health care (e.g. physician offices, clinics) facility environment. ELECTROMAGNETIC ENVIRONMENT – GUIDELINES Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Page 48
Electromagnetic immunity Portable and mobile radio equipment should be used no closer to the BEMER-SET PRO and its power cables than the recommended safety distance calculated from the equation applicable to the frequency transmitter. EMISSION TESTS IEC 60601 TEST LEVEL...
Page 49
AM/FM radio stations and TV stations cannot be predicted theoretically with accuracy. To determine the electromagnetic environment of a stationary transmitter, a site survey should be considered. If the measured field strength in the location in which the BEMER-SET PRO is used exceeds the conformity level listed above, the BEMER-SET PRO should be monitored to verify normal operation. If abnormal performance is observed, additional measures might be necessary, such as readjusting the BEMER-SET PRO or changing to another location.
Within the framework of the warranty obligation, the manu- or to which modifications have been made. All warranties and facturer accepts no liability for direct or indirect incidental or guarantees shall become void if the BEMER products are not consequential damage. used as intended.
Page 51
In all other respects, claims asserted by the customer for damages and/or the reimbursement of expenses, regardless of the legal grounds – in particular due to a breach of duties arising from the contractual obligations or unlawful acts – are excluded. This does not apply to product liability claims or in cases of intent and gross negligence or involving loss of life or bodily injury, or if the claims are based on the breach of an obligation whose...