Hitachi UST-9135P Instruction Manual page 47

Puncture electronic convex probe
Table of Contents

Advertisement

High-level disinfection
General name
Hydrogen peroxide
Peracetic acid
Glutaraldehyde
Glutaraldehyde
Glutaral
Note: * indicates that the marked disinfectant is not applicable in Canada.
Warning
After disinfection, thoroughly rinse the probe with deionized water.
Residual disinfectant can cause an adverse reaction to the operator or patient.
Trade name
*
PERASAFE
TM
Practical liquid 1.62W/V%
Acecide
*
®
Solution 6%
®
WAVICIDE
-01 *
Solution 2.65%
STERANIOS *
Solution 2.0%
Cidex plus
®
Solution 3.4%
-41-
Manufacturer
ANTEC INTERNATIONAL
Saraya Co., Ltd.
Medical Chemical Corporation
Laboratoires ANIOS
ADVANCED STERILIZATION PRODUCTS
A Johnson & Johnson company
Division of Ethicon, Inc.
MN1-5315 Rev. 14
®