  • Page 1 DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75 BioProTT Clamp-On SL User Manual...
  • Page 2 GmbH. © Copyright 2022 em-tec GmbH Finning. —All rights reserved. Subject to Technical Changes Owing to our policy of continuous product development, the illustrations and technical data contained in this document may differ slightly from the current version of the device.
  • Page 3: Table Of Contents

    DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75 Table of Contents System Description ......................4 Description of the Measurement Principle ..............5 General Safety Information, Symbols, Units and Abbreviations......6 Symbols Used in these Operating Instructions ..........6 3.2 Symbols on the BioProTT Clamp-On SL and on Packaging ......6 General Information Concerning this User Manual ...........
  • Page 4: System Description

    DocuSign Envelope ID: BE88571C-E493-4DFB-AF9A-864EB3FC3D75 BioProTT™ Clamp-On SL System Description em-tec's sensors, which are used in combination with, and powered by, one of the evaluation devices within the BioProTT Product Range, carry out measurements based on the ultrasonic transit time method. They enclose flexible tubes via a simple click-fastening and operate non-invasively, i.e., without physical contact...
  • Page 5: Description Of The Measurement Principle

    T and T , we are able to down determine the flow velocity. The calculation as such is carried out by the evaluation device. Copyright em-tec GmbH D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Page 5...
  • Page 6: General Safety Information, Symbols, Units And Abbreviations

    The device as a whole as well as any parts must be disposed of in accordance with WEEE Directive and national legislation. Manufacturer: em-tec GmbH · Lerchenberg 20 · 86923 Finning · Germany Year of manufacture Serial number Order number Temperature limit during storage and transport.
  • Page 7: General Information Concerning This User Manual

    For more information regarding applications the BioProTT Clamp-On SL can be used in, please contact em-tec GmbH. Copyright em-tec GmbH D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Page 7...
  • Page 8: Usage Restrictions And Limitations

    The user is responsible to use, check and maintain the BioProTT Clamp-On SL according to this user manual. em-tec GmbH is neither liable nor responsible for any consequences arising from the use of the BioProTT Clamp-On SL that does not comply with the user manual or specifications in this document.
  • Page 9: Maintenance And Service

    Clamp-On carried em-tec GmbH only. If these instructions are not followed, em-tec GmbH shall accept no liability for the component and the warranty will be void. If you experience any trouble with the measurement despite following the user manual, or if your BioProTT Clamp-On SL is damaged in any way, please contact our service department.
  • Page 10: Use Of The Bioprott Tm Clamp-On Sl

    Once this is done, the sensor can be attached to the tube. Please note: total cable length between respective evaluation device BioProTT Clamp-On SL shall not exceed 4 m. Page 10 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 11: Attaching The Bioprott

    For more information regarding the setup and handling of sensor and tube, refer to our TechNote Issue #1 "How to Optimize the Accuracy in Your Flow Measurement Application". Copyright em-tec GmbH D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Page 11...
  • Page 12: Use Of The Bioprott

    Before each measurement, the cables, connectors and pins must be checked for breaks or damages as this could result in wrong measurements. Do not use the BioProTT Clamp-On SL if you detect any damages. Page 12 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 13 It must be ensured that the correct calibration parameters are used. • While it is possible to carry out measurements outside of the specified flow range, em-tec GmbH cannot guarantee that resulting flow values will be within the given accuracy range.
  • Page 14: Disassembly And Storage Of The Bioprott

    Clamp-On SL is not suited for cleaning processes using machines. Sterilization processes, especially steam sterilization or autoclaving, may not be used. The following surface disinfectant is recommended: Name Manufacturer Contact Bacillol® AF Hartmann www.hartmann.de Page 14 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 15: Exchanging The Bioprott Tm Clamp-On Sl

    Ensure that the calibration table used is the one needed for your respective application. Optional Mounting Solutions In order to mount the sensor or to fix it in one position, em-tec offers optional mounting solutions. The features differ slightly in regard to their design and handling depending on the sensor size.
  • Page 16 If the mounting feature and thus sensor is not positioned correctly in regard to the tubing, e.g. due to kinking or excessive bending, this might impact the flow measurement and lead to incorrect flow values. For more information, see chapter 10.3 "BioProTT Mounting Solutions". Page 16 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 17: Adjustment And Calibration Of The Bioprott

    If any of these parameters change, it is possible to send the sensors in for a re-adjustment. In addition, em-tec GmbH recommends regular calibrations to be carried out. Ideally, the time between re-calibrations should not exceed 24 months. Other than the adjustment, the calibration can be carried out by the customer on site.
  • Page 18: Troubleshooting

    BioProTT™ Clamp-On SL Troubleshooting If there are any issues with the BioProTT Clamp-On SL, try the following suggestions. If the problem cannot be solved, please contact your local distributor or em-tec GmbH directly. Problem Possible Cause Action Check if the sensor is damaged.
  • Page 19 The connected sensor If the sensor is correct, dis- and then reconnect it. is the wrong sensor type. If the problem persists, return the sensor to em-tec GmbH for recalibration*. Sensor is not Check if the sensor is damaged or not connected recognized.
  • Page 20 Check if the selected calibration table is the one appropriate for your application. Wrong calibration If necessary, select the right calibration table. table selected. If the problem persists, contact our service department*. Page 20 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 21: Environmental Protection And Disposal

    In accordance with the requirements of EU Directive 2012/65/EC Waste Electrical and Electronic Equipment (WEEE), our customers in the EU are entitled to return all waste deriving from the products to us – in clean and disinfected condition. The em-tec GmbH WEEE registration number is: DE 99135207.
  • Page 22: Technical Specifications

    Clamp- On SL as well as the positioning of pumps and valves within – the circuit impact the measurement and must be taken into account when it comes to the accuracy of the BioProTT FlowMeasurement System. Page 22 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 23 *** ID = Inner Diameter; WT = Wall Thickness; OD = Outer Diameter Please specify the adjustment parameters at time of ordering, e.g. medium type, medium temperature, and tubing type (i.e. size and material). Copyright em-tec GmbH D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Page 23...
  • Page 24: Bioprott Tm Mounting Solutions

    Atmospheric Pressure 70 kPa to 106 kPa Temperature Range 10 °C to 40 °C (50 °F to 104 °F) Relative Humidity 10 % to 96 % (non-condensing) Page 24 D162-704 BioProTT™ Clamp-On SL – User Manual– V4.0 Copyright em-tec GmbH...
  • Page 25 BioProTT™ Clamp-On SL About em-tec GmbH em-tec has been a specialist for flow measurement systems in the medical and bioprocessing technology sector for over 30 years. The company's core competence is the non-invasive flow measurement using the ultrasonic transit-time method, that is used for applications in extracorporeal circulation systems of life- sustaining systems as well as in biopharma applications that use flexible tubes.