HYDAC ELECTRONIC AquaSensor AS 3000 Operation Manual

HYDAC ELECTRONIC AquaSensor AS 3000 Operation Manual

Hide thumbs Also See for AquaSensor AS 3000:
Table of Contents

Advertisement

Quick Links

 
 
 
 
 
 
 
 
 
 
 
 
 
 
Bedienungsanleitung
(Originalanleitung)
Operation manual
(Translation of original
instructions)
 
 
 
 
 
 
AquaSensor
AS 3000
Mit IO-Link Schnittstelle
With IO-Link Interface

Advertisement

Chapters

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AquaSensor AS 3000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for HYDAC ELECTRONIC AquaSensor AS 3000

  • Page 1       AquaSensor         AS 3000     Mit IO-Link Schnittstelle   With IO-Link Interface           Bedienungsanleitung (Originalanleitung) Operation manual (Translation of original instructions)          ...
  • Page 2: Table Of Contents

    Bedienelemente des AS 3000 Betriebsarten 11.1 SIO-Mode 11.2 SDCI-Mode Parametrierung 12.1 Allgemeines 12.2 Parametrierung im SDCI-Mode 12.3 Parametrierung im SIO Mode 12.4 Parametrierung mit HYDAC Programmiergerät HPG P1-000 12.5 Parametrierung mit HYDAC Programmieradapter ZBE P1-000 Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 3 Rücksetzen der Spitzenwerte Programmierfreigabe 18.1 Allgemeines 18.2 Ändern der Programmfreigabe Fehlermeldungen Anschlussbelegung Technische Daten Typenschlüssel Geräteabmessungen Zubehör 24.1 Elektrischer Anschluss 24.2 Mechanischer Anschluss Abkürzungsverzeichnis Kontakt Anhang / Annex: IODD Einstellparameter / IODD Setting Parameters Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 4 Gerätetechnik dem Zeitpunkt der Literaturerstellung entsprechen. Abweichungen bei technischen Angaben, Abbildungen und Maßen sind deshalb möglich. Entdecken Sie beim Lesen dieser Dokumentation Fehler oder haben weitere Anregungen und Hinweise, so wenden Sie sich bitte an: HYDAC ELECTRONIC GMBH Technische Dokumentation Hauptstraße 27 66128 Saarbrücken -Deutschland-...
  • Page 5: Sicherheitshinweis

    Das Symbol bedeutet, dass Tod, ein schwerer Personen- oder ein erheblicher Sachschaden eintreten können, wenn entsprechenden Vorsichtsmaßnahmen nicht getroffen werden. WARNUNG ! Symbol bedeutet, dass ein leichter Personen- oder Sachschaden eintreten kann, wenn entsprechenden Vorsichtsmaßnahmen nicht getroffen werden. VORSICHT ! Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 6: Bestimmungsgemäßer Gebrauch / Vorhersehbarer Fehlgebrauch

    100 und IEC 364 einzusehen (Bezugsquellen z.B. Beuth Verlag GmbH, VDE-Verlag GmbH).  Klare Regelung Verantwortlichkeiten für Montage, Installation, Inbetriebnahme und Bedienung sind festzulegen. Es besteht Beaufsichtigungspflicht bei zu schulendem oder anzulernendem Personal! Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 7: Organisatorische Maßnahmen

    Schaltmoleküle aus dem Fluid, was zu einer Änderung der Kapazität des Sensorelements führt. Der hieraus erhaltene Messwert stellt den Sättigungsgrad des Fluids in Prozent dar. Ein auf dem Sensor integriertes Thermoelement misst die Temperatur im Bereich von -25 °C bis +100 °C. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 8: Funktionen

    Es sollten deshalb alle sinnvollen Maßnahmen unternommen werden, um den Sättigungsgrad so niedrig wie möglich zu halten. Hydraulik- und Schmierfluide können nicht „zu trocken“ sein! Sättigungsgrad Sättigungs- 100% punkt 75 % 50 % 25 % Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 9: Wartung Und Instandhaltung

    6-3 und EN 61000-6-4 werden erfüllt. Die Forderungen der Normen werden nur bei ordnungsgemäßer und fachmännischer Erdung des Sensorgehäuses erreicht. Beim Einschrauben in einen Hydraulikblock ist es ausreichend, wenn der Block über das Hydrauliksystem geerdet ist. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 10: Zusätzliche Montagehinweise

    Auswahl des gewünschten Menü-Punktes und zum Einstellen der Werte. mode  Im Menü absteigen  Menüpunkt auswählen  Im Menü aufsteigen  Wert verkleinern  Wert bestätigen  Wert vergrößern Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 11: Betriebsarten

    Temperatur – Schaltbit 3 Boolean BDC3 Temperatur – Schaltbit 4 Boolean BDC4 14-bit Integer PDV2 Prozesswert 2 = Temperatur Nachfolgende Abbildungen zeigen die Anordnung und Reihenfolge der Bits und Octets für die Ausführung mit Temperaturfühler: Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 12: Parametrierung

    Der AquaSensor kann im SDCI-Mode über die IO-Link-Schnittstelle mit jedem IO-Link- fähigen Master-Konfigurationstool (gemäß IO-Spezifikation V1.1) parametriert werden. Die IO-Link-Spezifikation V1.0 wird ebenfalls unterstützt. Wird der eingelesene Parametersatz vom Gerät nicht akzeptiert, empfehlen wir den Parametersatz auf Plausibilität zu überprüfen. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 13: Parametrierung Im Sio Mode

    12.4 Parametrierung mit HYDAC Programmiergerät HPG P1-000 (Anschluss mit Standardkabel) HPG P1 Device Q1 / C 12.5 Parametrierung mit HYDAC Programmieradapter ZBE P1-000 (Anschluss mit Standardkabel) Q2 / QA ZBE P1 Device Q1 / C Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 14: Darstellung Der Digitalanzeige

    Ausgang 2 DS 1 Schaltverzögerungszeit, Ausgang 1 Schaltverzögerungszeit, Ausgang 2 Rückschaltverzögerungszeit, DR 1 Ausgang 1 Rückschaltverzögerungszeit, Ausgang 2 OV 1 Ausgang 1 Ausgang 2 Stromausgang Spannungsausgang Schließer bei Fensterfunktion Öffner bei Fensterfunktion Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 15 Max-Wert für Sättigungsgrad Min-Wert für Sättigungsgrad K IM Max-Wert für Temperatur Min-Wert für Temperatur Fehleranzeige Rücksetzen Nein RS.XL Reset Min / Max-Wert rS.HL Programmiersperre Tasten verriegelt FREE Tasten entriegelt Free UERS Version Vers Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 16: Ausgangsverhalten

    Messwert Abkürzungen: "SP1", "SP2" = Schaltpunkt 1 bzw. 2 "RP1", "RP2" = Rückschaltpunkt 1 bzw. 2 Eine Einstellung des Schaltpunktes (SP) ist nur möglich, wenn er oberhalb des zugehörigen Rückschaltpunktes (RP) liegt. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 17: Window-Mode - Einstellung Auf Fensterfunktion (Fno / Fnc)

    Schaltwertes (FL) liegt.  Die Fensterfunktion arbeitet nur dann ordnungsgemäß (Ein- und Aus- schalten), wenn alle Schaltwerte (inklusive Sicherheitszone) größer als 0 cm + Offsetwert und kleiner als die Messbereichsobergrenze + Offsetwert liegen. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 18: Einstellbereiche Für Die Schaltausgänge

    Messbereich) einstellen und den Messgrößen Sättigungsgrad bzw. Temperatur zuordnen. HINWEIS: Wird der Sensor an einem IO-Link-Master betrieben und Pin 2 als Analogausgang verwendet, dann darf dieser nicht mit dem IO-Link Master verbunden werden, da dieser keine Analogsignale unterstützt. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 19: Grundeinstellungen

    Zur Anpassung an die jeweilige Applikation kann das Verhalten des AS 3000 über mehrere Grundeinstellungen verändert werden. Diese sind zu einem Menü zusammengefasst. 15.1 Hauptmenü Der Bereich wird nicht angezeigt, wenn in „Erweiterte Funktionen“ unter „ou2“ „i“ oder „u“ eingestellt ist. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 20: Erweiterte Funktionen

    AS 3000 IO-Link 15.2 Erweiterte Funktionen Der Bereich wird nicht angezeigt, wenn in „Erweiterte Funktionen“ unter „ou2“ „i“ oder „u“ eingestellt ist. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 21 AS 3000 IO-Link Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 22: Ändern Der Grundeinstellungen

     "mode"-Taste betätigen.  Hier kann zwischen Programmierung frei „FREE“ und Programmierung gesperrt „Loc“ entschieden werden mit "" bzw. ““ erfolgt der Wechsel zwischen den Möglichkeiten.  Einstellung mit der "mode"-Taste quittieren. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 23: Fehlermeldungen

    Sie sich bitte mit unserer Service-Abteilung in Verbindung. HINWEIS:  Direkte Nähe zu Verbindungsleitungen von Leistungsverbrauchern oder störenden Elektro- oder Elektronikgeräten ist möglichst zu vermeiden  Bei Verwendung von geschirmten Kabeln mögliche Masseschleifen / Potentialverschleppungen vermeiden. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 24: Anschlussbelegung

    AS 3000 IO-Link Anschlussbelegung M12x1, 4-polig Q1/C Q2/QA Signal Bezeichnung Q2/QA Schaltausgang (SP2) / Analogausgang Q1/C Schaltausgang (SP1) / IO-Link Kommunikation Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 25: Technische Daten

    145 g Anmerkung: Verpolungsschutz , Kurzschlussfestigkeit sind vorhanden. FS (Full Scale) = bezogen auf den vollen Messbereich -25 °C mit FKM oder EPDM Dichtung, -40 °C auf Anfrage bei montierter Kupplungsdose entsprechender Schutzart Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 26: Typenschlüssel

    = Hydraulikflüssigkeiten auf Mineralölbasis = Hydraulikflüssigkeiten auf Phosphatesther-Basis Anschlussart mechanisch = G3/8 A ISO 1179-2 Anschlussart elektrisch = Gerätestecker M12x1, 4-pol. (ohne Kupplungsdose) Ausgang F31 = IO-Link Schnittstelle Modifikationsnummer 000 = Standard Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 27: Geräteabmessungen

    AS 3000 IO-Link Geräteabmessungen  2.11) 4-pol. Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 28: Zubehör

    5 m (197 inch) Leitung Material-Nr.: 6006789 Farbkennung: Pin 1: braun Pin 2: weiß Pin 3: blau Pin 4: schwarz 24.2 Mechanischer Anschluss ZBM 22 Adapter zum Einbringen in eine Leitung G1/2“ Bestell-Nr.: 3248511 Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 29: Abkürzungsverzeichnis

    Wert-Fenster oberer Wert Wert-Fenster unterer Wert Fensterfunktion Öffner Fensterfunktion Schließer Full Scale IODD IO Device Description ISDU Indexed Service Data Unit Ausgangsfunktion Rückschaltpunkt SDCI Single-drop digital communication interface Standard IO-Modus Schaltpunkt Schaltsignalkanal Stand: 14.07.2022 HYDAC ELECTRONIC GMBH Mat. Nr.: 670006...
  • Page 30: Kontakt

    AS 3000 IO-Link Kontakt HYDAC ELECTRONIC GMBH Hauptstr. 27 D-66128 Saarbrücken Germany Web: www.hydac.com E-Mail: electronic@hydac.com Tel.: +49 (0)6897 509-01 Fax.: +49 (0)6897 509-1726 HYDAC Service Für Fragen zu Reparaturen steht Ihnen der HYDAC Service zur Verfügung. HYDAC SYSTEMS & SERVICES GMBH Hauptstr.
  • Page 31 AquaSensor AS 3000 With IO-Link Interface Operating Instructions (Translation of original instructions)
  • Page 32 SDCI-Mode Parameterisation 12.1 General 12.2 Parameterisation in SDCI mode 12.3 Parameterisation in SIO Mode 12.4 Parameterisation with HYDAC programming device HPG P1-000 12.5 Parameterisation by means of HYDAC programming adapter ZBE P1-000 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 33 Change program enable Error Messages PIN connection Technical Data Model Code Device dimensions Zubehör 25.1 Electrical connection 25.2 Mechanical connection List of abbreviations Contact Anhang / Annex: IODD Einstellparameter / IODD Setting Parameters Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 34 Modifications to technical specifications, illustrations and dimensions are therefore possible. Should you find any errors whilst using these instructions, or have any suggestions for improvements, please contact: HYDAC ELECTRONIC GMBH Technical documentation Hauptstrasse 27 66128 Saarbruecken -Germany-...
  • Page 35: Safety Information

    WARNING! The symbol means that minor personal injuries or damage to property can occur if the stated precautions are not met. CAUTION ! Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 36: Intended Use / Foreseeable Misuse

    VDE 0105-100 and IEC 364 (source: e.g. Beuth Verlag GmbH, VDE-Verlag GmbH).  Clear regulation of responsibilities for the assembly, installation, start-up and operation to be defined. It is obligatory to provide supervision for trainee personnel! Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 37: Organisational Measures

    The value obtained represents the degree of saturation of the fluid as a percentage. A thermocouple built into the sensor measures the temperature in the range -25 .. +100°C. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 38: Functions

    There is no such thing as "too little water" in hydraulic and lubrication fluids! Saturation level Saturation 100% point 75 % 50 % 25 % Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 39: Maintenance And Servicing

    61000-6-3 and EN 61000-6-4. However, the requirements of those standards are met only if the sensor's housing has been correctly earthed by a qualified electrician. When fitted into a hydraulic block, earthing the block via the hydraulic system is sufficient. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 40: Additional Installation Instructions

    • One step back in menu • Choose menu item • Ascend in the menu • Reduce value • Confirm value • Increase value Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 41: Operation Modes

    Temperature – Switching bit 4 14-bit Integer PDV2 Process value 2 = Temperature The figures below show the arrangement and the order of the bits and octets for the version with a temperature sensor: Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 42: Parameterisation

    IO-Link compatible master configuration tool (according IO specifications V1.1). IO-Link specifications V1.0 are as well supported. Should the read parameter sets from the device not be accepted, we recommend to carry out a plausibility check of the parameter set. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 43: Parameterisation In Sio Mode

    (Connection via standard cable) HPG P1 Device Q1 / C 12.5 Parameterisation by means of HYDAC programming adapter ZBE P1-000 (Connection via standard cable) Q2 / QA ZBE P1 Device Q1 / C Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 44: Reading The Digital Display

    Switch-back delay time, output 2 Output 1 OV 1 Output 2 Current output Voltage output N/O when window function is active N/C when window function is active Unit conversion: temperature VN IM uniT Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 45 Min value for saturation level K IM Max value for temperature Min value for temperature Error indication Reset RS.XL Reset min/max value rS.HL Programming disable Keys locked FREE Keys unlocked Free UERS Version Vers Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 46: Output Signals

    = switch point 1 or 2 "RP1", "RP2" = switch-back point 1 or 2 It is only possible to set the switch point (SP) if it is higher than the respective switch-back point (RP). Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 47: Window Mode - Window Function Setting (Fno / Fnc)

    The window function only works properly (switching on and off) if all switch values (including the safety margin) are greater than 0 cm + offset value and below the upper measuring range limit + offset value. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 48: Setting Ranges For The Switching Outputs

    If the sensor is operated in conjunction with an IO-Link master and if pin 2 is used as an analogue output, pin 2 may not be connected with the IO-Link master, as it does not support analogue signals. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 49: Basic Settings

    The function of the AS 3000 can be adapted to suit the particular application by altering several basic settings. These settings are combined in a menu. 15.1 Main menu The range is not displayed if "i" or "u" is pre-set under "ou2" in the "extended functions". Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 50: Add-On Functions

    AS 3000 IO-Link 15.2 Add-on functions The range is not displayed if "i" or "u" is pre-set under "ou2" in the "extended functions". Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 51 AS 3000 IO-Link Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 52: Changing The Basic Settings

    Press "mode". • You can choose between free programming "FREE" and locked programming "Loc". By pressing "" or ““ you can switch between these options. • Confirm by pressing the "mode" key. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 53: Error Messages

    Keep the unit well away from the electrical supply lines of power equipment, as well as from any electrical or electronic equipment causing interference. • When using screened cables, possible ground loops / potential loss have to be avoided. Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 54: Pin Connection

    AS 3000 IO-Link PIN connection M12x1, 4 pole Q1/C Q2/QA Signal Designation Q2/QA Switching output (SP2) / analogue output Q1/C Switching output (SP1) / IO-Link Communication Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 55: Technical Data

    FS (Full Scale) = relative to complete measuring range -25 °C with FKM or EPDM seal, -40 °C on request with mounted mating connector in corresponding protection type Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 56: Model Code

    = Operating fluids based on phosphate esters Mechanical Connection = G3/8 A ISO 1179-2 Electrical connection = plug connector, M 12x1, 4 pole (mating connector not included) Output F31 =IO-Link interface Modification number 000 = Standard Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 57: Device Dimensions

    AS 3000 IO-Link Device dimensions Installation dimension ɸ 53,5 ∅ 2.11) ∅ 2.11) Connection M12x1 4 pole 4 pole Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 58: Zubehör

    Part No.: 6006789 Colour code: Pin 1: brown Pin 2: white Pin 3: blue Pin 4: black 25.2 Mechanical connection ZBM 22 Adapter for installing in a G1/2" line Order no.: 3248511 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 59: List Of Abbreviations

    FS Full Scale IODD IO Device Description ISDU Indexed Service Data Unit OU Output function RP Switch-back point SDCI Single-drop digital communication interface SIO Standard IO mode SP Switch point Switching signal channel Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 60: Contact

    AS 3000 IO-Link Contact HYDAC ELECTRONIC GMBH Hauptstrasse 27 D-66128 Saarbruecken Germany Web: www.hydac.com Email: electronic@hydac.com Phone: +49 (0)6897 509-01 Fax.: +49 (0)6897 509-1726 HYDAC Service For enquiries about repairs or alterations, please contact HYDAC Service. HYDAC SYSTEMS & SERVICES GMBH...
  • Page 61: Anhang / Annex: Iodd Einstellparameter / Iodd Setting Parameters

    AC%20ELECTRONIC%20GmbH%22) C%20ELECTRONIC%20GmbH%22) AS 3000 Family Version: V1.2 Release Date: 2018-05-28 AS 3006-F31-000 Vendor ID 348 (0x015c) Vendor Name HYDAC ELECTRONIC GMBH Vendor Text Hauptstr. 27, D-66128 Saarbrücken Vendor URL http://www.hydac.com Device ID 927224 (0x0e25f8) DeviceFamily AS 3000 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 62 7 - 0 15 - 8 7 - 0 Variable "BDC1 Behavior" index=61 id=V_Bdc1Behavior data type: 32-bit Record (subindex access not supported) access rights: rw modifies other variables excluded from data storage Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 63 UInteger Hysteresis octet bit offset 31 - 24 23 - 16 15 - 8 7 - 0 subindex element bit 7 - 0 7 - 0 15 - 8 7 - 0 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 64 31 - 24 23 - 16 15 - 8 7 - 0 subindex element bit 15 - 8 7 - 0 31 - 24 23 - 16 15 - 8 7 - 0 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 65 8-bit 3 = Two- V_Bdc3Behavior Mode UInteger Point 16-bit Lower -237..987 V_Bdc3Thresholds Integer Threshold 16-bit Upper -224..1000 V_Bdc3Thresholds Integer Threshold 32-bit 0..9999000 1000 On Delay UInteger 32-bit 0..9999000 1000 Off Delay UInteger Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 66 31 - 24 23 - 16 15 - 8 7 - 0 subindex element bit 15 - 8 7 - 0 31 - 24 23 - 16 15 - 8 7 - 0 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 67 0 = 4..20mA, 8-bit UInteger Analog Mode 1 = 0..10V octet bit offset 23 - 16 15 - 8 7 - 0 subindex element bit 7 - 0 7 - 0 7 - 0 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 68 31 - 24 23 - 16 15 - 8 subindex element bit 15 - 8 7 - 0 15 - 8 7 - 0 7 - 0 octet bit offset 7 - 0 subindex Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 69 Octet 2 bit offset subindex / / / / / / / / / / / / / / / / / / / / / / / / Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 70 / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 71 / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / / Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 72 Setpoint SP2 octet bit offset 31 - 24 23 - 16 15 - 8 7 - 0 subindex element bit 15 - 8 7 - 0 15 - 8 7 - 0 Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 73 8-bit 0 = Default, 1 V_SecondaryValueSp2 Switch Point Logic UInteger = Inverted 2 = Window, 8-bit 3 = Two- V_SecondaryValueSp2 Switch Point Mode UInteger Point 16-bit Switch Point V_SecondaryValueSp2 UInteger Hysteresis Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 74 129 (0x81) 17 (0x11) Parameter could not be saved to persistent memory saved Parameter could not be Parameter could not be loaded or restored from persistent 129 (0x81) 18 (0x12) loaded memory Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 75 Secondary value out Secondary value is below or above its valid range and limited to 36033 (0x8cc1) Warning of range the corresponding lower or upper value Status: 2022-07-14 HYDAC ELECTRONIC GMBH Part no.: 670006...
  • Page 76 Notizen / Notes / Notes                                                                        ...