4.9
Installation of the Inserts
Screw
the
chosen
handpiece or extension till the end of the
threading;
WARNING: Screw onto the
extension only the suitable inserts
supplied in the blister pack. Do not
screw any other type of inserts.
WARNING: Do not screw the insert
to be adopted with the extension
directly on the handpiece.
WARNING: PIEZOSURGERY®
medical+ inserts can only be installed
onto PIEZOSuRGERY® medical+
handpieces. PIEZOSURGERY® medical
inserts can only be installed onto
PIEZOSURGERY® medical handpieces.
Fasten the insert using the PIEZOSURGERY
medical or PIEZOSURGERY
torque wrench.
To use the Mectron torque wrench
correctly, proceed as follows:
Place the insert into the wrench as shown;
CAUTION: Do not use the
PIEZOSuRGERY® medical L torque
wrench to fasten the inserts. The
PIEZOSuRGERY® medical L torque
wrench must only be used to fasten the
extension.
INSTALLATION AND CONFIGuRATION
insert
into
the
®
medical+
®
1
2
3
19
Need help?
Do you have a question about the PIEZOSURGERY plus and is the answer not in the manual?