LITHOSTAR MODULARIS Error List

Displays on board D3

Once the system is turned on, different symbols are displayed chronologically on the dis-
play of board D3.
Display Possible cause/action
0
d
1
4
5
7
-

Text on the Console

"Chip card invalid"

This message is displayed when:
- the gold card is still logged on a different LITHOSTAR MODULARIS
- board D3 was replaced and the "init counter" on the gold card is already set at 3.
- the card is defective.
In normal mode, check whether the LED 174 on board D3 blinks temporarily and then
remains lit when turning it on.

LITHOSTAR MODULARIS SIEMENS

This message is displayed when too many errors occur. In this case, use the service PC
to generate an error log.

Missing Iso-center cross on the ultrasound system

There is no communication between the ultrasound system and the LITHOSTAR
MODULARIS.
Siemens AG
Medical Solutions
System o.k.
(only with the ultrasound localization option and if a communication was set
up to the ultrasound device.)
Software download.
Data transfer, the decimal point is flashing.
The receiving file is being checked.
Software download time too long.
Everything o.k.
(only if no ultrasound localization option is available or if the ultrasound device
is turned off.)
The service switch is set to normal mode - only for boards that still have no
software.
SPL1-130.840.02
Rev. 03
Page 13 of 14
04.05
CS PS 24
2 - 13
2
2
2
2
2
MODULARIS Uro Plus