Field
Description
22
Pressure support setting
23 to 29
Unused
30
Measured total
respiratory rate
31
Measured tidal volume
32
Measured total minute
volume
33
Unused
34
Measured peak
inhalation pressure
35 to 37
Unused
38
High inhalation pressure
alarm setting
39
Low inhalation pressure
alarm setting
40
Unused
41
Low exhaled mandatory
tidal volume alarm
setting
42
Low exhaled minute
volume alarm setting
43
High respiratory rate
alarm setting
1047358 Rev N
Table B-2: SNDA record format (continued)
Example
Resolution
0◆◆◆◆◆
1
◆◆◆◆◆◆
N/A
0.0◆◆◆
0.1 for 1.0
to 9.9
1 for 10 to
100
0.00◆◆
0.01
0.00◆◆
0.01 for
0.00 to
9.99
0.1 for 10.0
to 99.9
◆◆◆◆◆◆
N/A
50.0◆◆
0.1
◆◆◆◆◆◆
N/A
20◆◆◆◆
1
3◆◆◆◆◆
1
◆◆◆◆◆◆
N/A
0.00◆◆
0.01
0.0◆◆◆
0.1
0◆◆◆◆◆
1
Respironics V60 Ventilator User Manual
Communications interface
Range
Units
Comments
0 to 56
cmH
O
IPAP - EPAP in S/T
2
and PCV modes
0 in CPAP mode
"◆◆◆◆◆◆" in
AVAPS mode
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0.0 to 9.9
BPM
"◆◆◆◆◆◆" in
standby mode or
during exhalation
10 to 100
port leak test
0.00 to 9.99
L
"◆◆◆◆◆◆" in
standby mode, during
exhalation port leak
test, or for values out
of range
0.00 to 9.99
L
For values out of
range set output to
99.9◆◆
10.0 to 99.9
"◆◆◆◆◆◆" in
standby mode or
during exhalation
port leak test
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0.0 to 99.0
cmH
O
"◆◆◆◆◆◆" in
2
standby mode or
during exhalation
port leak test
N/A
N/A
Always output as
"◆◆◆◆◆◆"
10 to 50
cmH
O
2
0 to 40
cmH
O
2
N/A
N/A
Always output as
"◆◆◆◆◆◆"
0.00 to 1.50
L
Lo V
0.0 to 99.0
L
5 to 90
BPM
Appendix B
alarm setting
T
B-11