Chapter 8
Alarms and messages
Table 8-4: Vent Inoperative alarm messages: summary and troubleshooting (continued)
Message
Description
Vent Inoperative
Technical failure. The venti-
1009
lator is in the ventilator inop-
erative state.
Pressure Regula-
tion High
Vent Inoperative
Technical failure. The venti-
100A
lator is in the ventilator inop-
erative state.
Data Acquisition
PCBA ADC Refer-
ence Failed
Vent Inoperative
Technical failure. The venti-
100B
lator is in the ventilator inop-
erative state.
Watchdog Test
Failed
8-16
Corrective Action
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Check the patient. Provide al-
ternative ventilation. Have the
ventilator serviced.
Respironics V60 Ventilator User Manual
Priority
Manually
Autore-
type (ID)
resettable
settable
High
Yes
No
(11)
High
Yes
No
(12)
High
Yes
No
(13)
1047358 Rev N