Safety and performance
Guidance and manufacturer's declaration-electromagnetic immunity
EN
Withings Body Scan is intended for use in the electromagnetic environment specified below.
FR
The customer or the user of the Withings Body Scan should ensure that it is used in such an environment.
DE
Immunity test
SV
NL
Conducted RF
FI
IEC 61000-4-6
DA
Radiated RF
IT
IEC 61000-4-3
ES
CS
PL
Electrostatic discharge (ESD)
PT
IEC 61000-4-2
RO
Electrical fast transient/burst
HU
IEC 61000-4-8
SK
ET
Surge IEC 61000-4-5
EL
IS
Voltage dips, short
LV
interruptions and voltage
variations on power supply
LT
input lines IEC 61000-4-11
SL
TR
Power frequency (50/60Hz)
BG
magnetic field IEC 61000-4-8
HR
IEC 60601 Test Level
Compliance Level
3 Vrms
N/A
150kHz to 80MHz
3 V/m
N/A
80 MHz to 2.5 GHz
Contact: ±8 kV
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
Air: ±2 kV, ±4 kV, ±8 kV, ±15 kV
2 kV for power supply lines
N/A
1 kV for input/output lines
1 kV differential mode
line(s) to line(s):
2 kV common mode
+/- 1kV
-5% UT (95% dip in UT) for 0.5
N/A
cycles, -40% UT (60% dip in UT)
for 5 cycles, -70% UT (30% dip in
UT) for 25 cycles, -5% UT (95%
dip in UT) for 5 sec
30 A/m
30 A/m
50 Hz or 60 Hz
50 Hz or 60 Hz
Electromagnetic environment – guidance
N/A
Portable and mobile RF communications equipment should be used no
closer to any part of the equipment or system including cables than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity of
equipment marked with the following symbol:
Floors should be wood, concrete, or ceramic tile. If floors are covered
with synthetic material the relative humidity should be at least 30%.
The main power quality should be of the kind used in a typical commercial or
hospital environment.
The main power quality should be of the kind used in a typical commercial or
hospital environment.
The main power quality should be of the kind used in a typical commercial
or hospital environment. If the user of the equipment or system requires
continued operation during power main interruptions, it is recommended
that the equipment or system be powered from an uninterruptible power
supply or a battery.
The power frequency magnetic field should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
.