Error Messages

Error Messages

Series 50 Family

Message
Display
Err 9
US Toco
Err 14
US Toco
Err 16
US Toco
*

8040A

Message
Display
Err 16
US Toco
*. No longer available.

The following error messages are directly related to telemetry and appear on
the fetal monitor. Refer to the Instructions for Use provided with your monitor
for error messages not related to telemetry monitoring.
Cause
Invalid telemetry mode.
Incorrect transducer
connected to transmitter.
Transducers are
connected to the front
panel of the monitor.
Cause
Transducers are
connected to the front
panel of the monitor.
Solution
Check the cable from the telemetry
receiver and, if necessary, replace it.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Solution
Disconnect the transducers from the monitor or
switch off the telemetry receiver.
Chapter 3 - Tests and Error Messages