Decg Transducer (M1357A); Decg/Mecg Patient Module (M1364A); Iup Quartz Transducer (1290C #J05) - Philips 50 T Series Service Manual

Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

DECG Transducer (M1357A)

DECG/MECG Patient Module (M1364A)

IUP Quartz Transducer (1290C #J05)

Chapter 2 - Technical Specifications
Input Impedance:
CMRR:
Noise:
Contact Potential Tolerance:_500mV.
Input Voltage Range:
Patient Leakage Current:
Patient Auxiliary Current: <0.1_A (dc).
Dielectric Strength:
Weight:
Cable Length:
The patient module has a7-pin ECG connector into which you can plug
either DECG cable (M1362A or M1362B) or MECG cable.
Overall length:
Length of free cable:
Weight:
Size:
Socket:
A short cable version of the M1364A patient module with a 70 cm (28 inch)
cable is orderable as part number M1364-60003.
Dynamic Range:
Sensitivity:
Non-linearity:
Volume Displacement:
Weight:
Cable Length:
Length:
Height:
Operating Temp:
IUP Quartz Transducer (1290C #J05)
>10M (di_erential, dc to 50/60Hz).
>110dB (with patient cable, 51.5k/0.047_F
imbalance atline frequency).
<4_Vp (referred to input with 25k).
20_Vp to 3mVp.
<10_Arms @ 120V/60Hz.
2000Vrms (spark-gap protected).
185g (6.5oz).
2.5m (8ft 2in) or 70cm (28in) for Option C03.
2706mm (+30, -100mm)
2618mm (+30, -100mm)
120 grams
88x42x30mm
DECG or MECG connection
-50 to 300mmHg.
5_V/V/mmHg.
whichever is greater:
±1% of the reading OR
±1mmHg.
0.2mm3/100mmHg.
180g (6.3oz).
3m (9ft 10in).
37mm (1.5in).
21.7mm (0.9in).
15 to 40_C.