Outputs; Input Sensitivity; Output Power; Image Rejection - Philips 50 T Series Service Manual

Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

Outputs

Input Sensitivity

Output Power

Image Rejection

Self-Test Facilities

Chapter 2 - Technical Specifications
Transmitter
Cardio socket for M1356A ultrasound transducer or the M1357A/
M1364A DECG transducer.
Toco socket for the M1355A Toco transducer or an IUP Transducer
(CPJ840J5).
Remote Event Marker and Service Socket.
Receiver
25 pin RS232 Socket for connecting the Fetal Monitor.
-118dBM
Transmitter
4 mW (USA)
2 mW (Europe)
1 mW (Japan)
>80 dB
Self-test facilities include:
Receiver Self Test: Tests the Receiver LF Part.
Transmitter Self Test: Tests the Transmitter LF Part.
Parameter Test: With the appropriate transducer connected to the
Transmitter, the monitoring mode (ultrasound/DECG or Toco/IUP) is
tested from the Transmitter via Receiver to the connected Fetal
Monitor.
Telemetry System (Receiver and Transmitter)