AC INPUT POWER WIRING

AC POWER
115/230 VAC
1. Turn off power source.
2. Open terminal block cover.
3. Input Power connections:
a. Requires maximim of 2.4 watts
b. Connect AC1 to pin B1 and AC2 to pin B2
and AC Gnd to enclosure Gnd Lug
AC1
B1
B2
AC2
S A G E M E T E R I N G , I N C .
250–500Ω
C5
4. Connect 4-20 mA:
a. Connect isolated positive (+) lead to B4
b. Connect isolated negative (–) lead to C5
5. Connect a HART field calibrator as shown
insuring that there is a minimum of 250 ohm
resistance in the top loop.
6. For Pulsed output connect B4 (+) to C4 (–).
65
Loop
Power
+
Supply
B4
HART field
calibrator
REV. 10-SIX/SRX