O p e r a t i o n s a n d I n s t r u c t i o n M a n u a l
Approvals

HAZARDOUS LOCATION APPROVALS

Sage Rio Gas Flow Meter Type SIX or SRX are ATEX
approved.
Certification No.: TÜV 12 ATEX 7167 X
II 2 G Ex d IIB+H 2 T6 Gb
T
=–40˚F (–40˚C) to +122˚F (50˚C)
amb
Only use Silicone O-Rings for –40˚C which are
certified under DEMKO 07 ATEX 0622294U
UL Certification No. 20141219 - E464988
USL, CNL - Class I, Div 1; Groups B, C, and D
Hazardous Locations
CONFORMANCE
EN 60079-0:2009
EN 60079-1:2007
REQUIREMENTS
The cable entry holes shall be fitted with suitably
certified cable glands or suitably certified stopping
plugs that are capable of maintaining the IP rating
of the equipment.
• Do not open when energized.
• Do not open when an explosive atmosphere
might be present.
II 2 G Ex d IIB+H 2 T6 Gb
Equipment
group
II Non-Mining
EU Explosive
atmosphere symbol
Gas
Equipment
Type of Protection
category
Flameproof
2 = High protection
S A G E M E T E R I N G , I N C .
• Customer repair of the product (or replacement
of components) is not allowed.
• If the equipment is used in a manner not specified
by the manufacturer, the protection provided by
the equipment may be impaired.
• Permanently connected equipment shall employ a
switch or circuit breaker as a means for disconnec-
tion in the installment.
• Switch or circuit breaker must be suitably located
and easily reached.
• Switch or circuit breaker must be marked as the
disconnecting device for the equipment.
• The internal grounding terminal shall be used as
the equipment grounding means and that the
external terminal is only a supplemental bonding
connection where local authorities permit or
require such a connection.
TECHNICAL DATA FOR SAGE RIO MODELS SIX/SRX
BRIDGE CIRCUIT
24 VDC
I max
100 mA
P max
2.4 watts
I nominal
64 mA
P nominal
1.5 watts
TOTAL METER POWER
24 VDC
I max
100 mA
P max
2.4 watts
I nominal
64 mA
P nominal
1.5 watts
Gas Groups
IIB = Ethylene
+H2 = Hydrogen
31
110 VAC 50/60Hz 230 VAC 50/60 Hz
100 mA
100 mA
2.4 watts
2.4 watts
64 mA
64 mA
1.5 watts
1.5 watts
110 VAC
230 VAC
88 mA
44 mA
10.1 watts
10.1 watts
54 mA
27 mA
6.2 watts
6.2 watts
Temperature
Code
Max Surface Temp in
Gas
Zone I
Enclosure = 85˚C
REV. 10-SIX/SRX