Welch Allyn Atlas Service Manual page 28

Hide thumbs Also See for Atlas:
Table of Contents

Advertisement

Section 3 - Troubleshooting
MENU TESTS
2 WELCH ALLYN ATLAS MONITOR
3.1 Functional Test and Initial Diagnostic
Plug AC back in.
27. Press <DATE/TIME><TREND>
28. Get Advanced Configuration menu.
29. Write down all settings for resetting to customer prefer-
ence later.
30. Set language to your native tongue if necessary to
allow you to write down the other settings. The top
item is always the language, press either <SET> button
to step through list.
31. Press <PRINT> if printer fitted.
32. Press <TREND> to return to idle screen.
33. Press <DATE/TIME>
34. Get date/time menu.
35. Verify date/time, set if necessary.
35.1.
Bad date may indicate battery problem. If
date was bad, turn off unit, pull power cable,
wait 5 minutes reconnect power cable, turn
on unit. Check date again.
35.1.1.
Press Date/Time to return to waveform screen if needed.
36. Press <DATE/TIME><LEAD SELECT>
37. Get Service Mode menu.
38. Examine version/configuration data in lower half of
menu, and write it all down.
39. Press <SELECT> to highlight Reset to factory de-
faults.
40. Press <BP START/CANCEL> to reset configuration.
41. Press <TREND> to return to idle screen.
42. Press <DATE/TIME><TREND>
43. Get Advanced Configuration menu.
44. Set language to your native tongue if necessary
44.1.
We have just reset to factory defaults.
Compare settings to factory defaults appropri-
ate for the country – in Operator manual. If not
matching, indicates memory problems.
Changing only the language should not change
any of the other factory default settings.
45. Press <DATE/TIME><LEAD SELECT>
If date comes back bad: Replace
main board Model 200. Model 210,
220, if battery not dead replace main
board.