LiNX® Control System
Drive inhibit mode is indicated by a right-to-left chase sequence on the battery gauge.
The chase sequence continues until the fault condition is cleared.
8.4 Cut-off Voltage
When the battery voltage decreases below the battery cut-off voltage:
The status indicator flashes red (flash code 2, refer to 1.1 Fault codes and diagnosis codes, page 1)
l
The red LED on the battery gauge flashes
l
The horn sounds once every 10 seconds
l
29