Specifications

Model No.
Dimension
Weight
Power Source
Battery Life
Auto Turn-off
Test Sample
Sample Size
Test Time
Measurement Range
Meter/Strip Storage
Condition
Operating Condition
Control Solution
Storage Condition
Memory
The device has been certified to meet the electrical and safety
requirements of: EN/IEC 61010-1, EN/IEC 61010-2-101, EN/IEC
61326-1,EN/ IEC 61326-2-6, EN/IEC 60601-1-2 and EN/IEC
60601-1.

Statements

The equipment complies with IEC 61326-1. Emissions are low
and not likely to cause interference in nearby electronic
equipment.
The equipment is tested for immunity to electrostatic discharge
as specified in IEC 61000-4-2, at a level of 8kV for Air Discharge
and 6kV for contact discharge.
The equipment is tested for immunity to radio frequency
interference at the frequency range of 80 MHz to 2.7GHz.
BMB-EV099X
60mmx85mmx22.2mm
60g without batteries
Two 1.5V AAA alkaline batteries
Over 1000 tests
2 minutes after idle.
Capillary whole blood
0.3 microliter
5 seconds
20 to 600 mg/dL (1.1 to 33.3 mmol/L)
39.2°F to 104°F (
15~85% R.H.(noncondensing)
50°F to 104°F
(10°C to 40°C)
15~85% R.H.(noncondensing)
35°F to 86°F (2°C to 30°C)
400 blood glucose tests with date,
time and species indicator; species-
specific 7, 14 and 28 day averages
32
4°C to 40°C)
,
,