Note:
DO NOT squeeze the towel. It might cause meter
malfunction by soaking the test strip port with excess
liquid disinfectant.
Wash hands thoroughly with soap and water after handling
the meter, lancing device, or test strips.
Any trouble performing cleaning and disinfection procedure
on the meter and lancing device, please contact Customer
Support for technical assistance.
Stop using the meter or lancing device if signs of deteriora-
tion are noticed and contact customer service 24/7 at
1-866-373-2824 for help.
The CaviWipesXL has been shown to be safe for use with
PetTest BMB-EV099X blood glucose meter and lancing
device for 3,000 cleaning and disinfection cycles (6,000
times wiping).
Based on the robustness study performed by the
manufacturer, 3,000 cleaning and disinfection cycles
supports 1 to 2 cleaning and disinfection cycles per day
over the 5 years use life of the meter and lancing device.
The meter and lancing device must be replaced with a new
one at the end of cleaning and disinfection cycles or over
5 years use life, whichever comes first.
Prevent any liquid from entering the
test strip port!