Do you have a question about the HDP10000 Series and is the answer not in the manual?

Questions and answers

Yeasin khan
February 3, 2025

its not shwing actual voltage

User image 67a0953ca814b
1 comments:
Mr. Anderson
February 10, 2025

The Hantek HDP10000 Series may not be showing the actual voltage due to several possible reasons:

1. Load Regulation Issues: If the load regulation is not stable (≤0.1% + 5mV), it may cause voltage deviations.
2. Linear Regulation Variations: Small fluctuations in input voltage can affect the output voltage (≤0.01% + 5mV).
3. Readback Accuracy: The displayed voltage has an accuracy of ≤0.1% + 1 digit, which may cause slight deviations from the actual output.
4. Overvoltage Protection (OVP): If OVP is triggered, the output may be limited or shut off.
5. Ripple and Noise: The ripple voltage (10-15mVrms) can cause minor variations in the displayed value.
6. Instrument Warm-Up: The manual specifies that stable operation is achieved after 30 minutes of use within 18℃ to 28℃.

Checking these factors can help determine why the actual voltage is not being displayed correctly.