2.1 Basic performance
Range: Measured range in mH
O: 0~2mH
O....200mH
O
2
2
2
Measured range in Bar: 0~0.2bar... .350bar
Pressure type: Gauge pressure-G; Absolute pressure-A; Sealed gauge
pressure -S.
Accuracy (based on standard and test reference conditions, including
linearity, hysteresis, repeatability):
±1%FS (0~2mH
O≤range<3.5mH
O,0~0.2bar≤range< 0.35bar).
2
2
±0.5%FS (3.5mH
O≤range≤200mH
O,0.35bar≤range≤350bar).
2
2
±0.25%FS (range ≥10mH
O, range ≥1bar).
2
Compensated Temperature:-10℃~60℃
Long-term stability: ±0.2%FS/year.
Overload: 2 times full scale pressure
Vibration: 20g, 20Hz~2000Hz
Shock: 20g, 11ms
Protection rating: IP68.
Operation Temperature:
MPM426WPC level transmitter: -30℃≤Ta≤80℃(PFA)
MPM426WPF level transmitter: -20℃≤Ta≤80℃(PUR) / -10℃≤Ta≤70℃
(PVC/PE)
Storage Temperature:
MPM426WPC level transmitter: -40℃~85℃(PFA)
MPM426WPF level transmitter: -30℃~85℃(PUR) / -20℃~85℃
(PVC/PE)
The corresponding relationship between pressure type and range is
shown in Table 1 below: