5.4. Error Codes - During normal operation, the DST6xx continually performs
diagnostic testing on the sensor lines. Errors are indicated by the
following error codes:
Error Code
HI
Reading is above meters usable range
LO
Reading is below meters usable range or sensor is shorted
CbL1
Indicates cable 1 is open (J1 pin 1)
CbL2
Indicates cable 2 is open (J1 pin 2)
CbL3
Indicates cable 3 is open (J1 pin 3)
CbL4
Indicates cable 4 is open (J1 pin 4), or open sensor
CbL5
Indicates a short between pins 2 & 3
err
Indicates accuracy error detected. Will also display during rapid
changes of temperature to the RTD sensor.
Prb
Indicates probe ID does not match probe ID programmed in meter
- - -
Indicates Low Battery Shutdown Mode
1°/0.1°Jumper
Battery Positive
J7 DST600R
Terminal
PCB
Ground Wire
DST600 Series USER MANUAL
Description
°F/°C Jumper
J8
Only
RTD +.
Exc-
Probe ID
Signal
RTD -
Probe ID
.
Ground
Fig. 5 (DST600R Shown)
9 of 20
Calibration
Pin 1
J1
Exc+
Port